Section 6-Cylinder Fill Times; Troubleshooting; Replacement Parts - Accessories - Invacare HomeFill HF2POST6 Operator's Manual

Oxygen post valve cylinder
Table of Contents

Advertisement

SECTION 6—CYLINDER FILL TIMES
SECTION 6—CYLINDER FILL TIMES
NOTE: All filling times are approximate and may vary on
environmental conditions.
*
Cylinder
Filling
Times
NORMAL
MAXIMUM

Troubleshooting

If post valve cylinder fails to function, try the following
solutions. If problem cannot be corrected, consult your
dealer/home care equipment provider.
PROBLEM
No flow.

Replacement Parts - Accessories

HF2POST6BAG or HF2POST9BAG Carrying Bag.
HomeFill
Concentrator Flow Rate to Patient:
(Platinum 5 up to 3 L/min. and Platinum 10 up to 5
HF2POST6
1 hr 25 min
1 hr 50 min
POSSIBLE CAUSE
1. Regulator or
conserver is in "Off"
position.
2. Cylinder empty.
3. Toggle Valve closed.
4. Conserver not
sensing breath.
L/min)
HF2POST9
2 hr 20 min
2 hr 45 min
1. Set to prescribed
setting.
2. Replace/fill cylinder.
3. Open toggle valve.
4a. Check position of
cannula in nose. Do not
breathe through mouth.
4b. Check to see if the
cannula is kinked.
20
HF2POSTD
3 hr 20 min
3 hr 50 min
SOLUTION
Part No 1148114

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Homefill hf2post9Homefill hf2postd

Table of Contents