Filling The Cylinder - Invacare HomeFill HF2POST6 Operator's Manual

Oxygen post valve cylinder
Table of Contents

Advertisement

SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
2. Ensure toggle valve is open. Open the toggle valve by
rotating counterclockwise.
3. Examine the gauge on the connected conserver or
regulator. Perform one of the following:
• If the cylinder pressure is less than 1500 psig, fill the
cylinder. Connect the cylinder to the compressor. Refer
to Filling the Cylinder on page 13.
• If the cylinder pressure is greater than 1500 psig., DO
NOT attempt to fill this cylinder. It may not fill. Use
cylinder as normal.

Filling the Cylinder

NOTE: For this procedure, refer to FIGURE 4.2 thru FIGURE 4.7.
1. Check the cylinder pressure before attempting to fill
cylinder. Refer to Checking Cylinder Pressure on page 12.
2. Close the toggle valve by rotating clockwise (FIGURE 4.2).
3. Remove the conserver or regulator according to the device
manufacturer instructions (FIGURE 4.2).