Invacare HF2RE9ES User Manual

Invacare HF2RE9ES User Manual

1.7l (m9) cylinder with continuous flow regulator for homefill oxygen system

Advertisement

Quick Links

2014/68/EU
93/42/EEC
This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.
Invacare 1.7L (M9) Cylinder With
Continuous Flow Regulator
For HomeFill® Oxygen System
en HF2RE9ES
User Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HF2RE9ES and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Invacare HF2RE9ES

  • Page 1 Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator For HomeFill® Oxygen System en HF2RE9ES User Manual 2014/68/EU 93/42/EEC This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
  • Page 2 All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
  • Page 3: Table Of Contents

    Contents 6 After Use ........24 6.1 Disposal .
  • Page 4: General

    Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator 1 General Read the manual 1.1 Symbols No open flame Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or DO NOT Smoke property damage.
  • Page 5: Intended Use

    General 1.2 Intended Use 1.3 User Profile This cylinder assembly is intended to store, regulate, and No specific product knowledge or training is required, other deliver supplemental oxygen to patients who may have than what is contained in this manual, to operate the difficulty extracting oxygen from the air they breathe.
  • Page 6: Warranty

    Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator 1.5 Warranty Terms and conditions of the warranty are part of the general terms and conditions particular to the individual countries in which this product is sold. 1.6 Component Identification Item Description...
  • Page 7: Safety

    Safety 2 Safety 2.1 Label Locations Serial number located above product label shown here. 1167426-E...
  • Page 8: General Guidelines

    – This cylinder is to be used with the Invacare equipment. HomeFill oxygen system. This cylinder – If you are unable to understand the warnings, contains 93±3% oxygen produced by an...
  • Page 9 – DO NOT SMOKE while using this product. – DO NOT use near OPEN FLAME or IGNITION SOURCES. – DO NOT use any lubricants on the cylinder unless recommended by Invacare. – NO SMOKING signs should be prominently displayed. – Avoid creation of any spark near oxygen equipment.
  • Page 10 Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator Operating Information DANGER! Risk of Fire DANGER! Dirt or contaminants on the cylinder assembly or Risk of Fire regulator can cause fire, injury or death. Dirt or contaminants on the cylinder assembly or –...
  • Page 11 – DO NOT use if particulates are inhaled. – DO NOT immerse product in any kind of liquid. – Contact Invacare or dealer for service. – DO NOT allow any kind of liquid to get inside openings of the regulator.
  • Page 12 Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator CAUTION! DANGER! Risk of Injury or Damage Risk of Injury or Death To reduce the risk of injury or damage from The cannula can cause tripping, falling, or other dropped or impacted cylinder assemblies: injury if improperly positioned and secured.
  • Page 13 The biocompatibility of cylinder accessories may vary by manufacturer. ALWAYS confirm prescribed dose before administering to Invacare accessories such as the cylinder bag and patient and monitor on a frequent basis. Invacare oxygen tubing are latex and phthalate free.
  • Page 14: Setup

    Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator 3.3 Receiving Inspection 3 Setup Remove product from package and inspect for damage. 3.1 Dealer Setup If there is any damage, DO NOT use. Contact your dealer/Home Care equipment provider. WARNING! – Before the first patient use, the provider MUST 3.4 Cylinder Prefill Inspection...
  • Page 15: The Hydrostatic Testing Date

    Setup DANGER! DANGER! Risk of Injury or Death Risk of Death, Injury or Damage from Fire Improper filling or abuse of this cylinder assembly, To avoid fire, death, injury or damage: or failure to follow these safety instructions – Cylinders that have been refinished and/or may cause serious bodily injury or death.
  • Page 16: External Examination

    Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator 3.4.2 External Examination WARNING! Risk of Injury or Damage To reduce the risk of injury or product damage: – DO NOT use the cylinder if any of the following items are found. Contact your dealer/Home Care equipment provider for replacement.
  • Page 17: Usage

    – All cylinders MUST be inspected before attempting to fill. – If the cylinder is found to be defective or does not work properly, contact an Invacare dealer or qualified technician for service or replacement. Before filling a cylinder assembly, read and follow the...
  • Page 18 Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator 1. Prior to each use, inspect product for visible damage. 6. Attach a standard or high flow single lumen adult Refer to the External Examination section. DO NOT use oxygen nasal cannula with a maximum length of 2.1 m if any damage is found.
  • Page 19 10. Turn flow dial E until the prescribed setting is aligned with the flow indicator window F. WARNING! – To avoid injury to patient, ALWAYS confirm prescribed setting before administering to patient and monitor flow on a frequent basis. – Only use the Invacare Corporation carrying bag. 1167426-E...
  • Page 20: Maintenance

    Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator 5 Maintenance DANGER! Risk of Fire Dirt or contaminants on the cylinder assembly or 5.1 Maintenance regulator can cause fire. – DO NOT use or fill cylinder assembly if the WARNING! cylinder fillport is dirty or contaminated.
  • Page 21: Emptying The Cylinder Assembly

    Maintenance 5.2 Emptying the Cylinder Assembly DANGER! Risk of Fire – NEVER smoke in an area where oxygen is being administered. – NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere. – DO NOT direct flow of oxygen at any person, or flammable material when adjusting the flow dial.
  • Page 22: Cleaning

    Invacare 1.7L (M9) Cylinder With Continuous Flow Regulator 5.4 Disinfecting 5. Examine the cylinder gauge H until the pressure is below 500 psig (3447 kPa). Refer to the safety information in Maintenance Cylinder pressure is below 500 psig (3447 kPa) before performing this procedure.
  • Page 23: Reuse

    Before transferring the product to a different Risk of Injury or Damage user, providers MUST inspect the product for Use of non-Invacare accessories may result in damage and disinfect the product. Refer to the serious injury or damage. following procedures: –...
  • Page 24: After Use

    6.2 Wear and Tear Invacare reserves the right to ask for any item back that has an alleged defect in workmanship. See Warranty that shipped with the product for specific warranty information.
  • Page 25: Troubleshooting

    Troubleshooting 7 Troubleshooting 7.1 Troubleshooting If the Regulating Cylinder fails to function, consult the Troubleshooting guide. If problem cannot be corrected, consult your Dealer/Home Care equipment provider. Problem Possible Cause Solution No flow. 1. Flow dial at setting “0”. 1. Adjust flow dial to prescribed 2.
  • Page 26: Technical Data

    The cylinder assembly is filled with USP 93±3% oxygen using its proprietary fill port that allows 6.0 ± 0.60 filling only through the use of an Invacare HomeFill System. The duration of oxygen flow between fills is dependent on Pressures the size of the cylinder and the flow setting.
  • Page 27 Technical data Environmental Operating Temperature Range Platinum 9 Up to 5 L/min Operating Temperature -20°C to 60°C (–2°F to Range 140°F) All filling times are approximate and may vary on environmental conditions. 95% RH Max Operating Humidity Max Altitude 3071 m (10,000 ft) Accessories Storage Temperature Range –40°C to 60°C (-40°F to Single Lumen Cannula...
  • Page 28 Notes...
  • Page 29 Notes...
  • Page 30 Notes...
  • Page 31 Ireland • Tel: (353) 1 810 7084 • Fax: (353) 1 810 7085 • ireland@invacare.com • www.invacare.ie Italia: Invacare Mecc San s.r.l., Via dei Pini 62, I-36016 Thiene (VI) • Tel: (39) 0445 38 00 59 • Fax: (39) 0445 38 00 34 • italia@invacare.com • www.invacare.it Nederland: Invacare AE, Galvanistraat 14–3, NL–6716 BZ Ede •...
  • Page 32 Invacare Corporation EU Representative Invacare Deutschland GmbH Kleiststraße 49 One Invacare Way D-32457 Porta Westfalica Elyria, Ohio USA Germany 44035 Tel: (49) (0) 5731 754 0 440–329–6000 Fax: (49) (0) 5731 754 52191 800–333–6900 Technical Services 440–329–6593 800–832–4707 www.invacare.com Manufacturer Invacare Corp.