Invacare HomeFill HF2POST6 Operator's Manual page 15

Oxygen post valve cylinder
Table of Contents

Advertisement

SECTION 4—INSPECTION AND COMPRESSOR INTERFACE
NOTE: If the outer ring (sleeve) is in the UP position (GREEN
dots not visible), the connector fillport will not be able to
accept the cylinder fillport. Pushing DOWN momentarily
will reset the connector fillport (GREEN dots visible) to
accept the cylinder fillport.
PUSH
DOWN
GREEN Dots
FIGURE 4.4 Resetting Connector Fillport
6. Grasp the cylinder assembly as shown in Detail "A" in
FIGURE 4.5.
7. Position the cylinder in the compressor cradle as shown in
Detail "C" in FIGURE 4.5.
8. Align the cylinder fillport with the connector fillport as
shown in Detail "C" in FIGURE 4.5.
9. Pull UP on the outer ring (sleeve) of the connector fillport
(Detail "B") while pushing DOWN on the cylinder
assembly to couple the cylinder fillport into the connector
fillport as shown in Detail "B" in FIGURE 4.5.
NOTE: The cylinder is properly connected when an audible "click"
is heard.
10. Press the compressor power switch to the On ( ) position.
Part No 1148114
Outer Ring (Sleeve)
BEFORE
COUPLING
CYLINDER,
PUSH DOWN
ON SLEEVE
UNTIL
GREEN DOTS
ARE VISIBLE.
15
Connector Fillport
HomeFill