SECTION 7—CYLINDER FILL TIMES
SECTION 7—CYLINDER FILL TIMES
NOTE: All filling times are approximate and may vary on
environmental conditions.
*
Cylinder
Filling
Times
NORMAL
MAXIMUM

Troubleshooting

If Conserving Cylinder fails to function, consult the
Troubleshooting guide. If problem cannot be corrected,
consult your Dealer/Home Care equipment provider.
PROBLEM
No flow.
Oxygen Conserving Cylinder
Concentrator Flow Rate to Patient:
(Platinum 5 up to 3 L/min. and Platinum 10 up to 5
POSSIBLE CAUSE
1. Regulator in "Off"
position.
2. Cylinder empty.
3. Dial not set at
increment.
4. Conserver not
sensing breath.
L/min)
HF2PCM2KIT
25 min
30 min
1. Set to prescribed
setting.
2. Replace/fill cylinder.
3. Set dial so indictor
points to a setting.
4. Check position of
cannula in nose. Do not
breathe through mouth.
5. Check to see if the
cannula is kinked.
24
SOLUTION
Part No 1143247