Page 1
Invacare® 1.0L/1.7L Cylinder with Dual Lumen Conserver For HomeFill® Oxygen System en HF2ECE6, HF2ECE9 User Manual 2014/68/EU 93/42/EEC This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
Page 2
All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
Invacare® 1.0L/1.7L Cylinder with Dual Lumen Conserver 1 General Read the manual 1.1 Symbols No open flame Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or DO NOT Smoke property damage.
General 1.2 Intended Use product. It is not intended for supplying oxygen therapy during refilling. This product is intended to be used by or This cylinder is intended to store, regulate, and deliver under adult supervision only after reading and understanding supplemental oxygen to patients who may have difficulty the instructions and warnings of this user manual.
– This cylinder is to be used with the Invacare equipment. HomeFill oxygen system. This cylinder – If you are unable to understand the warnings, contains 93±3% oxygen produced by an...
Page 9
– DO NOT SMOKE while using this product. – DO NOT use near OPEN FLAME or IGNITION SOURCES. – DO NOT use any lubricants on the cylinder unless recommended by Invacare. – NO SMOKING signs should be prominently displayed. – Avoid creation of any spark near oxygen equipment.
Page 10
Invacare® 1.0L/1.7L Cylinder with Dual Lumen Conserver Operating Information DANGER! Risk of Fire DANGER! Dirt or contaminants on the cylinder assembly or Risk of Fire conserver can cause fire, injury or death. Dirt or contaminants on the cylinder assembly or –...
Page 11
– DO NOT use if particulates are inhaled. – DO NOT immerse product in any kind of liquid. – Contact Invacare or a dealer for service. – DO NOT allow any kind of liquid to get inside openings of the conserver.
Page 12
Invacare® 1.0L/1.7L Cylinder with Dual Lumen Conserver CAUTION! DANGER! Risk of Injury or Damage Risk of Injury or Death To reduce the risk of injury or damage from The cannula can cause tripping, falling, or other dropped or impacted cylinder assemblies: injury if improperly positioned and secured.
Page 13
The biocompatibility of cylinder accessories may vary by manufacturer. ALWAYS confirm prescribed dose before administering to Invacare accessories such as the cylinder bag and patient and monitor on a frequent basis. Invacare oxygen tubing are latex and phthalate free.
– Oxygen cylinders shall be used for personal Hydrostatic Testing Date section. use only. Not to be filled for resale or use by professional users. – Filling of cylinders is performed by Invacare WARNING! HomeFill oxygen system only. Risk of Injury –...
Setup DANGER! DANGER! Risk of Injury or Death Risk of Death, Injury or Damage from Fire Improper filling or abuse of this cylinder, or failure To avoid fire, death, injury or damage: to follow these safety instructions may cause – Cylinders that have been refinished and/or serious bodily injury or death.
Invacare® 1.0L/1.7L Cylinder with Dual Lumen Conserver 3.4.2 External Examination WARNING! Risk of Injury or Damage To reduce the risk of injury or product damage: – DO NOT use the cylinder if any of the following items are found. Contact your dealer/Home Care equipment provider for replacement.
– All cylinders MUST be inspected before attempting to fill. – If the cylinder is found to be defective or does not work properly, contact an Invacare dealer or qualified technician for service or replacement. Before filling a cylinder, read and follow the information in...
Page 18
Invacare® 1.0L/1.7L Cylinder with Dual Lumen Conserver 1. Prior to each use, inspect product for visible damage. 5. Attach a dual lumen adult oxygen nasal cannula with Refer to the External Examination section. DO NOT use a maximum length of 2.1 m (7 feet) to the product’s if any damage is found.
Page 19
• Continuous Flow L basis. • Conserving Pulse M – Only use the Invacare Corporation carrying The conserver is labeled “CONTINUOUS” and bag. “PULSE”. The first three flows (0.25, 0.50 and 0.75 L/min) are in continuous mode at all times, regardless of the switch position.
Page 20
Invacare® 1.0L/1.7L Cylinder with Dual Lumen Conserver 11. Breathe through the nose and feel a pulse of oxygen as you inhale. When on a pulse setting, there is flow, or a pulse, only at the beginning of each breath. If you do not feel the pulse at the beginning of each breath, check the setting.
– The cylinder MUST be disconnected from the on product. HomeFill and emptied of gas prior to cleaning – All repairs MUST be performed by Invacare or disinfecting. Refer to the Emptying the Corporation or in an Invacare approved repair Cylinder Assembly section.
Invacare® 1.0L/1.7L Cylinder with Dual Lumen Conserver 5.2 Emptying the Cylinder Assembly DANGER! Risk of Fire – NEVER smoke in an area where oxygen is being administered. – NEVER use near any type of flame or flammable/explosive substances, vapors or atmosphere.
Maintenance 6. Turn the flow dial D to setting “0”. Follow the cleaning/disinfectant product manufacturer’s 7. Close the shutoff valve I fully. instructions for use, handling, storage and disposal of the product. Turn the shutoff valve clockwise I to close. This procedure is for disinfecting the cylinder and is to be performed by the provider prior to a patient change and 5.3 Cleaning...
Invacare® 1.0L/1.7L Cylinder with Dual Lumen Conserver 5.6 Storage 1. Ensure the shutoff valve I is closed. 2. Ensure the flow dial D is turned to setting “0”. 3. Place the cover on the fillport H. 4. Store product in a clean, dry and cool area free from grease, oil, and other sources of contamination.
6.2 Wear and Tear Invacare reserves the right to ask for any item back that has an alleged defect in workmanship. See Warranty that shipped with the product for specific warranty information.
Invacare® 1.0L/1.7L Cylinder with Dual Lumen Conserver 7 Troubleshooting 7.1 Troubleshooting If Conserving Cylinder fails to function, consult the Troubleshooting guide. If problem cannot be corrected, consult your dealer/Home Care equipment provider. Problem Possible Cause Solution No flow. 1. Flow dial at setting “0”.
The cylinders are filled with USP 93±3% oxygen using their proprietary fillports that allow filling only through the use of an Invacare HomeFill System. The duration of oxygen flow between fills is dependent on the size of the cylinder and the flow setting.
Page 30
Invacare® 1.0L/1.7L Cylinder with Dual Lumen Conserver Environmental Operating Temperature Range Storage Temperature Range –20°C to 60°C (-4°F to 140°F) Operating Temperature Range 1.7°C to 40.6°C (35°F to 105°F) Max Operating Humidity 95% non-condensing Assembly Dimensions Model HF2ECE6 HF2ECE9 Diameter 11.1 cm (4.4 in)
Page 31
Technical data Fill Times Model HF2ECE6 HF2ECE9 1 hr 25 min* 2 h 20 min* * Concentrator flow rates to patient: Platinum 5, Platinum S Series and Perfecto Series Up to 2.5 L/min Platinum 9 Up to 5 L/min All filling times are approximate and may vary on environmental conditions.