3.2.1 System Status Lights (Leds)

The five red status indicators show the status of the Static Weight Indicator
NET
ON indicates when a net weight is displayed, in other words, it is on when a tare weight has
been acquired.
STABLE WEIGHT
The stable weight condition is determined on the basis of two parameters :
motion band and motion delay.
Motion band defines a range in which the weight is stable, and motion delay defines the time
the weight should stay in the range before stable weight indications turns on.
ZERO WEIGHT
Net weight is considered zero when its value is between +/- 0.5 scale division.
The weight must stay in this range for 2 seconds after the weight becomes stable zero weight
indication is turned on.
ALARM
Alarm indication flashes if an alarm is pending, either the alarm is NEW or has been
ACKNOLOWLEDGED.
READY
Ready indication turns on if the scale is calibrated (zero and span calibration complete) and no
SHUT DOWN conditions are active.

3.2.2 Keypad

The keypad is comprised of pad touch keys consisting of the following
F1
F2
F3
^
v
RUN / SETUP
On the interface scroll, the fourth display row is composed of three sections. Each section shows
the action of the corresponding key.
Baud Rate
> 9600 <
<F1 Descr.>
<F2 Descr.> <F3 Descr.>
The Scroll Keys (UP or DOWN) allow change the page displayed.
The RUN/SETUP Key allows selection of the menu modes. RUN is the operative mode with
indication of variables and status scale. SETUP is the configuration and calibration mode of the
instrument.
Function Key 1
Function Key 2
Function Key 3
Scroll Up
Scroll Down
Selection mode RUN / SETUP
3-2