Specifications - Fluke T+ Calibration Information Manual

Electrical tester
Hide thumbs Also See for T+:
Table of Contents

Advertisement

T+ and T+ PRO
Calibration Information

Specifications

Temperature Performance
Operating ........................................................... −10 °C to 55 °C (14 °F to 131 °F)
Storage ............................................................... −30 °C to 60 °C (−22 °F to 140 °F)
Relative Humidity
90 % .................................................................... 0 °C to 30 °C (32 °F to 86 °F)
75 % .................................................................... 30 °C to 40 °C (86 °F to 104 °F)
45 % .................................................................... 40 °C to 50 °C (104 °F to 122 °F)
Altitude
Operating ............................................................ 2,000 m
Storage ............................................................... 10,000 m
Battery Type/Life ................................................... AAA (2); 40 hours
Shock, Vibration .................................................... Sinusoidal vibration per MIL-PRF-28800F for a Class 2 instrument
Safety ..................................................................... ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010-1 to 1000 V CAT
AC Bandwidth ....................................................... 45 Hz to 66 Hz
Earth Ground ......................................................... MAXIMUM working voltage 1000 V ac or dc
Duty Cycle .............................................................. Indefinitely for voltages up to 240 V.
Voltage Hazard LED .............................................. LEDs turn on @ voltages > 30 V ac/dc ± 35 %
LEDs ....................................................................... LEDs turn on @ between 70 % and 100 % of the indicated voltage on
Accuracy
AC voltage .......................................................... ± (3 % rdg + 2 digits)
DC voltage .......................................................... ± (2 % rdg + 2 digits)
Resistance .......................................................... ± (5 % rdg + 3 digits)
LCD Resolution ................................................... 0.1 V for voltages < 50 V, 1 V for voltages ≥ 50 V, 0.01 kΩ at
GFCI Test Current .............................................. 100 V to 150 V @ 6 mA to 9 mA ac
Standard Input Test Current ................................ < 5 mA
Size (HxWxL) ......................................................... 1.3 x 2.15 x 7.6 in.
Weight .................................................................... 9.9 oz
4
III/600 V CAT IV
MAXIMUM measurable voltage 600 V ac or dc
This will be displayed on the LED and LCD (T+ PRO)
For voltages between 240 V to 600 V, the duty cycle is 30 s on / 300 s
off. For voltages above 240 V, the Tester must connect to a voltage
source only for a MAXIMUM of 30 s and then disconnect for a
MINIMUM of 300 s.
the LED, except for the 12 V LED that turns on @ between 50 % and
100 %.
resistance measurement
150 V to 240 V < 12 mA