Advertisement
To prevent possible electrical shock, fire, or personal injury:
The Calibration Manual for the 1550C/1555Insulation Tester (hereafter referred to as "the product") provides the following information.
For complete operating instructions and additional safety information, refer to the 1550C/1555Users Manual located on the CD that came with the product.
To contact Fluke, call one of the following telephone numbers:
To register your product, visit http://register.fluke.com.
To see, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals.
A Warning identifies conditions and actions that pose hazard(s) to the user; A Caution identifies conditions and procedures that could cause Product damage, equipment under test damage, or permanent loss of data.
Read before product use.
To prevent possible electrical shock, fire, or personal injury:
Symbols on the product and in the manual are explained Table 1.
Table 1. Symbols
Symbol | Meaning |
Risk of Danger. Important information. See Manual. | |
Hazardous voltage | |
![]() | Equipment protected by double or reinforced insulation. |
![]() | Do not use in distribution systems with voltages higher than 1100 V. |
![]() | Interference is present. Displayed value might be outside of specified accuracy. |
![]() | Ramp mode indicator |
![]() | Electrical breakdown |
![]() | Volts AC |
![]() | Earth Ground |
Equipment required to perform the procedures in this manual is listed in Table 2. If the recommended models are not available, equipment with equivalent specifications may be substituted.
For safe operation and maintenance of the product, have an approved technician repair the product.
Do not attempt to use the 5500A, 5520A, or other standard calibrator for insulation and continuity resistance tests. Calibrator damage will result.
Table 2. Required Equipment
Equipment | Minimum Required Characteristics | Recommended Model |
HV Probe | 6 kV, ±1% (1000:1 Divider) 11 kV, ±2% for the 1555 | Fluke 80K-6 80K-15 |
Digital Multimeter | 500 mVdc to 1 V: ±0.02% | Fluke 8508 |
1Load with Guard Terminal | Resistances 200 kΩ, ±1.25%, 500 V 500 kΩ, ±1.25%, 500 V 1 MΩ, ±1.25%, 1 kV 2.5 MΩ, ±1.25%, 2.5 kV 5 MΩ, ±1.25%, 5 kV 10 MΩ, ±1.25%, 10 kV 1 GΩ, ±1.25%, 10 kV 100 GΩ, ±1.25%, 10 kV 200 GΩ, ±1.25%, 10 kV 500 GΩ, ±5%, 10 kV 1 TΩ, ±5%, 10 kV 2 TΩ, ±5%, 10 kV | Combinations of: Welwyn F Series, Welwyn MFP2 Series And Vishay HTS-523 |
2 Capacitors w/ Bleeder Resistors | 0.1 µF, ±5%, 500 V, Polypropylene 1 µF, ±5%, 2.5 kV, Polypropylene | |
Calibrator | DC current: 2 mA Accuracy: ±1.25 % DC Voltage: 0 - 550 V Accuracy: ±0.005 % AC Voltage: 0 - 240 V, 60 Hz Accuracy: ±1.25% | Fluke 5080, Fluke 5520A |
IR Cable Assembly | Fluke P/N 2166275 | |
3 Calibration Software | Snorre | |
Ammeter | Fluke 8508 | |
Personal computer | IBM compatible, with Microsoft Windows XP SP2 or later +.NetFramework 2.0 or later | |
1550C/1555Users Manual CD ROM | Fluke P/N 3592810 |
To prevent possible electrical shock, fire, or personal injury, do not contact the output terminals while performing the following procedures. There are potentially dangerous voltages at the output terminals when the product is in the MΩ TEST function.
The following performance tests should be completed yearly to ensure that the product, referred to as "the UUT" (Unit Under Test) in this section of the manual, is in proper operating condition and meets the published accuracy specifications. If the UUT fails any of the performance test steps, repair or adjustment is needed. Refer to How to Contact Fluke for service information.
IR Port Verification Test
To verify operation of the IR Communications Port:
Button Test
Use the pushbuttons to control the product, view test results, and scroll through chosen test results. Pushbuttons and their functionality are discussed in Figure 1.
Figure 1. Pushbuttons
Item | Description |
1 | Turns the product off and on. |
2 | Push ![]() |
3 | Scrolls through test voltages, stored test results, timer duration, and changes test tag ID characters. Also used to answer "yes" to yes/no prompts. |
4 | After a memory location is set, ![]() |
5 | Use to scroll through test voltages, stored test results, timer duration, and memory locations. Also used to answer "no" to yes/no prompts. |
6 | Use for Test Voltage mode to start incrementally setting the test voltage between 250 V and 10,000 V. |
7 | Starts and stops a test. Push and hold for 1 second to start a test. Push again to stop a test. |
Display Test
Turn the UUT on several times while observing the display during power up. Compare the display with the example in Figure 2. Check all segments for clarity and contrast.
Figure 2. Display Features
Item | Description |
1 | Interference present. Readings could be out of specified accuracy range. |
2 | Polarization Index |
3 | Dielectric Absorption Ratio |
4 | Electrical breakdown in Ramp mode |
5 | Ramp mode indicator |
6 | Possible hazardous voltage is at the test terminals. To prevent personal injury, before and after testing, confirm that the product does not indicate the presence of hazardous voltage. If the product beeps continuously and there is hazardous voltage, disconnect test leads and remove power from the circuit under test. |
7 | Voltage sourced by the product or from the circuit under test at terminals of the product. |
8 | Test voltage selection (250 V, 500 V, 1000 V, 2500 V, 5000 V, or 10,000 V) |
9 | Battery charge status |
10 | Bar graph display of insulation resistance |
11 | Digital display of insulation resistance |
12 | Text display. Shows voltage, test current, capacitance, programmable test voltages, and menu options. |
Charging Test
Note
A fully-charged battery is indicated when the battery symbol shows all segments. Recharge the battery as necessary to obtain all segments. A full charge may require 12 hours.
Insulation Accuracy Test
Using the various resistances shown in Table 3, perform the UUT insulation accuracy test. Push for 2 seconds to start or discontinue a test.
Notes
The capacitance reading is obtained by pressing after a test has started.
Table 3. Insulation Accuracy Test
Step | Voltage Range | Resistance | UUT Display Limits | |
Minimum | Maximum | |||
1 | 250 V | 0.1 µF | 0.055 | 0.145 |
2 | 500 V | 250 kΩ | 237 kΩ | 263 kΩ |
3 | 500 V | 1 GΩ | 0.95 GΩ | 1.05 GΩ |
4 | 500 V | 100 GΩ | 80 GΩ | 120 GΩ |
5 | 1 kV | 1 GΩ | 0.95 GΩ | 1.05 GΩ |
*6 | 2.1 kV | 1 µF | 0.82 | 1.18 |
7 | 2.5 kV | 1 GΩ | 0.95 GΩ | 1.05 GΩ |
8 | 5 kV | 1 GΩ | 0.95 GΩ | 1.05 GΩ |
9 | 5 kV | 100 GΩ | 95 GΩ | 105 GΩ |
10 | 5 kV | 1 TΩ | 0.80 TΩ | 1.20 TΩ |
11 | 5 kV | 5 MΩ | 4.75 MΩ | 5.25 MΩ |
**12 | 10 kV | 1 GΩ | 0.95 GΩ | 1.05 GΩ |
**13 | 10 KV | 200 GΩ | 190 GΩ | 210 GΩ |
**14 | 10 KV | 2 TΩ | 1.6 TΩ | 2.4 TΩ |
**15 | 10 KV | 10 MΩ | 9.5 MΩ | 10.5 MΩ |
** 1555 only * Use "Programmable Test Voltage" mode by pushing ![]() |
Output Voltage Test
In Table 4, the UUT output voltage is checked with various loads applied. In this test a voltmeter with a high-voltage probe must be connected to the load resistor to measure the UUT output voltage. Use 15 80K-6 for voltages below 6 kV.
Table 4. Output Voltage Test
Step | Voltage Range | Load Resistor | Reading Limits | |
Minimum | Maximum | |||
1 | 250 V | 250 kΩ | 250 V | 275 V |
2 | 250 V | No Load | 250 V | 275 V |
3 | 500 V | 500 kΩ | 500 V | 550 V |
4 | 500 V | No Load | 500 V | 550 V |
5 | 1 kV | 1 MΩ | 1000 V | 1100 V |
6 | 1 kV | No Load | 1000 V | 1100 V |
7 | 2.5 kV | 2.5 MΩ | 2500 V | 2750 V |
8 | 2.5 kV | No Load | 2500 V | 2750 V |
9 | 5 kV | 5 MΩ | 5000 V | 5500 V |
10 | 5 kV | No Load | 5000 V | 5500 V |
*11 | 10 kV | No Load | 10000 V | 11000 V |
*12 | 10 kV | 10 MΩ | 10000 V | 11000 V |
* 1555 only. Requires 80K-15 probe. |
Short Circuit Current Test
To verify the UUT short circuit current, use the following procedure:
Table 5. Short Circuit Current Test
UUT | Voltage Range | Reading Limits | |
Minimum | Maximum | ||
1550C | 5000 V | 1.20 mA | 1.80 mA |
1555 | 10000 V | 1.20 mA | 1.80 mA |
Voltage Measurement Accuracy
To verify voltage measurement accuracy of the Live Circuit Warning function, apply the voltages listed in Table 6 to the + and - terminals of the UUT. Verify that:
Table 6. Voltage Measurement Test
Step | Voltage Source Output | UUT Display | UUT Tone | UUT Display Limits | |
Minimum | Maximum | ||||
1 | -34 Vdc | Flashing Hazard | Beeps | 29 V | 39 V |
2 | 240 Vac, 60 Hz | Flashing Hazard | Beeps | 204 V | 276 V |
The product should be performance tested yearly to ensure compliance with its specifications. When required, use the following adjustment procedure to bring the UUT within its nominal accuracy specifications.
Interface Connection
Perform adjustment with software using a computer and IR (infrared) adapter.
Connect the Infrared Cable Assembly to the UUT IR Port and COM port of the computer. Refer to Figure 3.
Figure 3. IR Port
Item | Description |
1 | IR Port |
2 | IR Device |
Instrument Setup
Turn the product on and wait for Test Voltage to appear on the display. From the computer terminal, activate the Snorre program from the Windows Start menu. On the Setup tab, confirm the selected COM port settings.
To prevent possible electrical shock, fire, or personal injury, do not contact the output terminals or test equipment terminals while performing the following procedures. There are potentially dangerous voltages when the UUT is in the "Calibrate HV Output and Measurement" mode.
Normalizing the HV Probe and Digital Multimeter
HV Adjustment
Current Adjustment
Charge Adjustment
Note
The following additional procedures are used during factory calibration and repair but should not be performed in the field. They are included for information only.
Various diagnostics are available from the Diagnostic tab as follows:
Identification (Id)
This button installs the default opvars in the UUT; the present calibration is lost as the opvars are overwritten with the defaults. The user is presented with a YES/NO dialog box, to prevent accidental selection of this option.
Restart UUT
This button first sets the UUT to CAL_DIAGS mode the sends out the restart hardware command. The UUT is then restarted.
Shutdown UUT
This button puts the UUT into CAL_DIAGS mode and sends out the powerdown command.
Get Diagnostics
Pressing this button continually gets Raw ADC values from the UUT and updates the Raw ADC Counts boxes (v_counts, i_counts, q_counts), pressing the button again turns this feature off.
Query Constants
Pressing this button provides an html screen dump of the present opvars obtained from the UUT. You will need to check that numerical values are reported for all nine variables. See Figure 4.
Figure 4. Query Constant Result
Save / Print html Page
The Save button brings up a Windows Save Dialog Box so that the html document being displayed can be saved to file.
The Print button brings up a Windows Print Dialog Box so that the rendered html document may be printed.
The window at the bottom of the page is a scrollable log of the Snorre methods issuing commands to the UUT and the corresponding responses received back from the UUT.
A Message From Fluke Corporation
Some semiconductors and custom IC's can be damaged by electrostatic discharge during handling. This notice explains how you can minimize the chances of destroying such devices by:
The following practices should be followed to minimize damage to S.S. (static sensitive) devices.
For safe operation and maintenance of the product:
To prevent possible damage to the product or to equipment under test:
Note
This instrument contains a lead-acid battery. Do not mix with the solid waste stream. Spent batteries should be disposed of by a qualified recycler or hazardous materials handler. Contact your authorized Fluke Service Center for disposal and recycling information.
Storing rechargeable lead-acid batteries in a low-charged state could lead to reduced life and/or damage. Fully charge the battery before storing for extended periods and check, the charge at regular intervals.
The product is powered by 12 V lead acid battery, Fluke P/N 2803592. The battery can be recharged using the AC power cord.
Fully charging the battery typically takes 12 hours. Avoid charging in extremes of temperature. Recharge the battery if the product has been stored for extended periods.
If the battery needs to be replaced, use the following procedure to replace the battery.
Disassembly
To prevent possible damage to the product or to equipment under test, disassembly must be performed using proper ESD handling techniques. Place the product on an anti-static mat and use a grounded wrist strap during the following procedure.
Re-assembly
For safe operation and maintenance of the product, remove excess water from the cloth before cleaning the product to ensure that water does not enter any terminal.
Periodically wipe the case with a damp cloth and mild detergent. Do not use abrasives or solvents to clean the product.
Replacement parts are listed in Table 7.
Table 7. Replacement Parts
Parts | Part Number |
Test Lead Set Fluke 1550 | 3477137 |
10 kV Clip | 3611951 |
1550C 1555 Safety Input Cover | 3529198 |
1550C Top Case | 3622602 |
1555 Top Case | 3624655 |
Input Jacks Decal | 3624643 |
Case Screws | 3552926 |
IR Cable Assembly | 2166275 |
Battery Hold Bracket | 3540654 |
Case Bottom | 3524293 |
Battery | 2803592 |
Rubber Foot | 3777953 |
AC Power Cord (S. Africa) | 1552363 |
AC Power Cord (Australia) | 658641 |
AC Power Cord (UK) | 769455 |
AC Power Cord (Continental Europe) | 789422 |
AC Power Cord (North America) | 284174 |
Soft Carrying Case | 3592805 |
Extended Lead Set (5 kV rating) | 2032761 |
1550C/1555 Quick Reference Card | 3592822 |
1550C/1555 CD ROM (includes Users Manuals [English, French, German, Spanish]) | 3592810 |
General Specifications
Display | 75 mm x 105 mm | |
Power | 12 V lead-acid rechargeable battery. 2.6 Ahr | |
Charger Input (AC) | 85 V to 250 V ac, 50/60 Hz, 20 VA This Class II (double insulated) instrument is supplied with a Class 1 (grounded) power cord. The protective earth terminal (ground pin) is not connected internally. The extra pin is for added plug retention only. | |
Dimensions (H x W x L) | 170 mm x 242 mm x 330 mm (6.7 in. x 9.5 in. x 13.0 in.) | |
Weight | 3.6 kg (7.94 lbs.) | |
Temperature (operating) | -20 °C to 50 °C (-4 °F to 122 °F) | |
Temperature (storage) | -20 °C to 65 °C (-4 °F to 149 °F) | |
Relative Humidity | 80% to 31 °C decreasing linearly to 50% at 50 °C | |
Altitude | 2000 m | |
Enclosure Sealing | IP40 | |
Input Overload Protection | 1000 V ac | |
Electromagnetic Compatibility | EN 61326-1, EN 61326-2-2 | |
Safety Compliance | EN 61010-1, EN 61557 Parts 1 and 2 CAT III 1000V, CAT IV 600V | |
Pollution Degree | 2 | |
Typical Battery Charge Capability Note: At temperature extremes, the battery needs to be charged more frequently. | Test Voltages | Number of Tests |
250 V | 4100 | |
500 V | 3600 | |
1 kV | 3200 | |
2.5 kV | 2500 | |
5 kV | 1000 | |
10 kV | 500 |
Electrical Specifications
Product accuracy is specified for 1 year after calibration at operating temperatures of 0°C to 35°C. For operating temperatures outside the range (-20°C to 0°C and 35°C to 50°C), add ±0.25% per °C, except on the 20% bands add ±1% per °C.
Insulation | |||
Test Voltage (DC) | Insulation Resistance Range | Accuracy (±reading) | |
250 V | <200 kΩ 200 kΩ to 5 GΩ 5 GΩ to 50 GΩ >50 GΩ | unspecified 5 % 20% unspecified | |
500 V | <200 kΩ 200 kΩ to 10 GΩ 10 GΩ to 100 GΩ >100 GΩ | unspecified 5 % 20% unspecified | |
1000 V | <200 kΩ 200 kΩ to 20 GΩ 20 GΩ to 200 GΩ >200 GΩ | unspecified 5 % 20% unspecified | |
2500 V | <200 kΩ 200 kΩ to 50 GΩ 50 GΩ to 500 GΩ >500 GΩ | unspecified 5 % 20% unspecified | |
5000 V | <200 kΩ 200 kΩ to 100 GΩ 100 GΩ to 1 TΩ >1 TΩ | unspecified 5 % 20% unspecified | |
10,000 V | <200 kΩ 200 kΩ to 200 GΩ 200 GΩ to 2 TΩ >2 TΩ | unspecified 5 % 20% unspecified | |
Bar graph range: Insulation test voltage accuracy: Induced ac mains current rejection: Charging rate for capacitive load: | 0 to 2 TΩ -0%, +10% at 1 mA load current 2 mA maximum 5 seconds per µF | ||
Discharge rate for capacitive load: | 1.5 s/µF |
Leakage Current Measurement | Range | Accuracy |
1 nA to 2 mA | ±(20% + 2 nA) | |
Capacitive Measurement | 0.01 µF to 20.00 µF | ±(15% of reading + 0.03 µF) |
Timer | Range | Resolution |
0 to 99 minutes | Setting: 1 minute Indication: 1 second | |
Live circuit warning | Warning Range | Voltage Accuracy |
30 V to 1100 V ac/dc, 50/60 Hz | ±(15% + 2 V) |
Short circuit current
> 1 mA and < 2 mA
The product measures insulation parameters and displays the results using with the following formulas.
Ohm's Law | Capacitance (charge) | PI (Polarization Index) | DAR (Dielectric absorption ratio) |
![]() | ![]() | ![]() | ![]() |
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Advertisement
Need help?
Do you have a question about the 1550C and is the answer not in the manual?
Questions and answers