2
Installation

Powering On the LC Pump for the First Time

Powering On the LC Pump for the First Time
28
Surveyor LC Pump Plus Hardware Manual
Figure 23. Outlet tubing connection to the purge manifold
Drain tubing
5. Place the free end of the drain tubing into a waste receptacle.
CAUTION To minimize solvent fumes, the waste receptacle should have a tight cap
with a hole for the drain tubing. You can also use this waste receptacle to collect the
drainage from the autosampler waste line and the detector outlet tubing.
After installing the solvent lines and tubing, you can power on the pump.
To power on the LC pump
1. Ensure that the power switch at the front of the LC pump is in the Off position (released
or out position).
2. Attach the power cord to the power entry module on the back panel of the pump and
connect it to line power.
3. Turn the power on by pushing the power button in to engage it.
The Power LED turns solid green. If the LED does not illuminate, see the first
troubleshooting tip in
Thermo Fisher Scientific service representative for assistance.
Table 8
on
page
98. If you cannot resolve the problem, call your
7/16 in. metal nut
Thermo Scientific