Adding And Programming Accessories - Culligan Premier Series Installation, Operation And Service Instructions

Automatic deionizer water treatment systems models from 2012
Table of Contents

Advertisement

Adding and Programming Accessories

TDS Probe
The standard Premier Series Auto DI ships with a TDS probe installed as shown in
is connected to the DI GBE controller board as shown in
is shown below. The user enters the "TDS Limit" which is the level of product water TDS which will, if exceeded for the
length of the quality delay set (see
quality rinse cycle.
In some applications, it is desirable to use a conductivity monitor instead of a TDS. In these case use the optional external
quality sensor (see
page
86), follow the directions on that page to properly configure the external sensor.
2)MANUAL REGEN
3)SET DATE/TIME
4)ADV. SETUP
>5)ACCESSORIES
>1)TDS
2)FLOW METER
TDS SETUP
>TDS PROBE
TDS LIMIT
>20
PPM
NOTE
If the TDS drops below the limit before MIN CAPACITY (GALS) is reached, a PURGE will occur. If
quality does not rise above the limit, a regeneration will be triggered. If MIN CAPACITY is set to zero,
the unit will trigger a regeneration immediately upon falling below the limit.
TDS ALARM
>200
PPM
TDS OFFSET
>0
PPM
NOTE
An external monitor may also be used to trigger regeneration. A monitor with a dry contact can be
used and wired to the Aux Input on the GBE board. When the dry contact closes, it will signal the de-
ionizer to regenerate. See
TEMP OFFSET
>0F
Culligan® Premier Series Automatic Deionizer
36
Figure 11 on page
"Auxiliary External Outputs" on page
1.
From the main menu screen, press
2.
Press
. The screen displays the Accessories menu.
3.
Press
to select 1)TDS
4.
The default configuration is a standard TDS probe measurement. Press
the TDS probe settings.
5.
The TDS probe monitors the total dissolved solids in the deionizer system. Set the
high TDS limit that will trigger a regeneration. To change the value, press
or
and
. The screen displays the next TDS setting, TDS ALARM.
6.
Press
and
or
the TDS exceeds the TDS alarm limit for a 15-second period the system will dis-
play error mode "Hi TDS Alarm."
7.
Press
when done. The screen displays the TDS limit setting.
8.
This setting is used to calibrate the TDS probe in the system. Measure the actual
TDS of the product water using a calibrated TDS meter, then press
to adjust the TDS offset so the displayed value matches the calibrated reading.
9.
Press
when done. The screen displays the TDS offset temperature setting.
"Conductivity Meter Quality Monitor" on page
10. The TDS Probe monitors the ambient temperature at the tank location. If the actual
location temperature is different than the probe reading, press
set the temperature offset value.
11. Press
when done. The menu returns to the Advanced Setup menu.
Figure 10 on page
26. The programming for the TDS probe
94), will cause the AutoDI unit to enter the
to 5)ACCESSORIES.
.
The screen displays the TDS Setup settings.
to adjust the TDS limit that will trigger a system alarm. If
86.
19. This probe
to view
and
and
or
and
or
to
Cat. No. 01024000