System Setup

Single System Setup
The Culligan® Premier Series Automatic Deionizer can be configured to operate with one or two deionizer systems.
1)INFORMATION
2)MANUAL REGEN
3)SET DATE/TIME
>4)ADV. SETUP
>1)SYSTEM SETUP
2)REGEN SETUP
3)CYCLE TIMES
4)AUXOUT SETUP
SYSTEM SETUP
>SINGLE
Twin Alternating Mode for Multi-System Setup
NOTE
If the system uses two sets of deionizer systems, the user-specified settings apply to both systems.
ALTERNATING MODE
>MASTER
PFLOW UPDATE
>UPDATE NOW
NOTE
PFLOW UPDATE works only in one direction. SLAVE1 deionizer settings can NOT be applied to the
MASTER deionizer.
NOTE
Only one system set should ever be in service at any given time.
Regeneration Trigger Setup
MINIMUM CAPACITY
>1000
Gals
CAPACITY
>100000
Gals
Culligan® Premier Series Automatic Deionizer
34
1.
From the main menu screen, press
2.
Press
. The screen displays the Advanced Setup menu.
.
3.
Press
The screen displays the System Setup settings.
4.
The default configuration is a single deionizer system. The deionization system is
available as a duplex alternating configuration. If you purchased a duplex system,
press
and
or
at SINGLE.
5.
If you selected an ALTERNATING configuration, you must set up each deionizer
in the system as either MASTER or SLAVE1; SLAVE1 must have identical settings
as the MASTER. Press
or SLAVE1.
6.
If you selected the MASTER deionizer in the system, you can press
select UPDATE NOW, which applies the MASTER device settings to SLAVE1.
7.
This setting is used to trigger a regeneration of the deionizer based on the capaci-
ty remaining (default = 1000 gallons). This setting requires the optional flow meter.
The batch setting begins with the Capacity setting (below) and counts down. When
the reading gets to this value a regeneration will be triggered. To change the value,
press
and
or
8.
This setting is the gallons to be treated between regenerations (default = 100,000
gallons). When this value reaches the minimum capacity, the system triggers re-
generation. This value is reset after the regeneration. To calculate the capacity,
use the CAAP® sizing program. The capacity of the Anion Resin normally dictates
the service run length of the deionizer. The capacity of the strong base anion resin
is determined by four factors: the regenerant level (lbs of sodium hydroxide per
ft
), and the content of Monovalent Ions, Carbonic Acid, and Silica in the raw wa-
3
ter. For example, at a Sodium Hydroxide dosage of 6 lbs per ft
ions, 10% Carbonic Acid, and 5% Silica, the Strong Base Anion Capacity is 16,000
grains per ft
. If Silica increases to 10%, the capacity decreases to 15,700 grains
3
per ft
. To change the value, press
3
to 4) ADV. SETUP.
to change the setting to ALTERNATING; otherwise leave
and
or
to change the setting to either MASTER
. Press
when done.
and
or
to
, 25% Monovalent
3
. Press
when done.