Appendix - Thermo Scientific STANDARD Series Manual

Laboratory temperature control products
Hide thumbs Also See for STANDARD Series:
Table of Contents

Advertisement

Thermo Scientific
Serial Communications

Appendix

1. Terminal settings
-
Baud rate:
-
Data bits:
-
Stop bits:
-
Parity:
-
Flow control:
-
Local echo:
2. General syntax
-
Read command:
Command:
o
Return on success:
o
Return on error:
o
-
Write command:
Command:
o
Return on success:
o
Return on error:
o
-
Remarks:
Parameter Id consists of 2 characters
o
Decimal point is a point "."
o
<SP>
o
<CR>
o
<LF>
o
Fxxx
o
-
Examples:
W HA 0 85.5
o
R LA 0
o
LA +10.00
19200
8
1
none
none
on
(the thermostat does not send a command echo)
R<SP><parameter Id><CR><LF>
<parameter Id><value><CR><LF>
Fxxx<CR><LF>
W<SP><parameter Id><SP><value><CR><LF>
$<CR><LF>
Fxxx<CR><LF>
Space character
Carriage return character
Line feed character
Failure code F000 to F255
Write parameter HA 0 with the value 85,5
Read parameter LA 0
LA 0 has a value of 10,00
A-1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents