Zimmer MIS Multi-Reference 4-in-1 Femoral Instrumentation Surgical Technique
Option 2
MIS QS Notch Guide
Position the appropriate size MIS QS
Notch Guide onto the femur so it is
flush against the resected surfaces
both distally and anteriorly. The MIS
QS Notch Guide will not contact the
anterior chamfer. Use the previously
prepared trochlear recess and/or the
femoral post holes to position the MIS
QS Notch Guide mediolaterally.
Fig. 4bb
Secure the MIS QS Notch Guide to
the femur with two 3.2mm (1/8-inch)
Headed Screws or predrill and insert
two 3.2mm (1/8-inch) Holding Pins
(Fig. 4bb). Use a reciprocating saw
to cut the sides and the base of the
intercondylar notch (Fig. 4cc). Then
remove the MIS QS Notch Guide (Fig.
4dd).
Fig. 4cc
Fig. 4dd
21