Option SD0 - Internal Storage Disabled Upgrade

Hardware Removal and Replacement

Outer Cover Removal
8. Turn the instrument off and disconnect the power cord.
9. Referring to
Figure
the strap handle.
Figure 3
Outer Cover Removal and Installation
10.Remove the two hole plugs from the opposite side of the instrument (not shown in
11.Remove the four rear feet (3) by removing the fours screws that attach them.
12.Remove the four bottom feet (4) by first removing the locking keys and then lifting the tabs
and sliding the foot toward the tabs.
10
3, remove the two screws (1) that attach the strap handle (2) and remove
Figure
Installation Note N5180-90087
3).