[:SENSe[1]:]LIST:TSLot:EXCLude:OFFSet:TIME
Syntax
[:SENSe[1]:]LIST:TSLot:EXCLude:OFFSet:TIME <numeric_value>
[:SENSe[1]:]LIST:TSLot:EXCLude:OFFSet:TIME?
Description
This command is used to set the exclusion area offset time relative to the beginning of the
time slot.
The query returns the current exclusion area offset time.
Parameter
Item
numeric_value
Example
LIST:TSL:EXCL:OFFS:TIME 0.01
Reset condition
On reset, the value is 0 s.
Error messages
– If DET:FUNC is set to AVER, error –221, "Settings conflict" occurs.
– If the limits of the values are exceeded, error –222, "Data out of range;value clipped to
upper (or lower) limit" occurs.
– If LIST:TSL:EXCL:OFFS:TIME is set while a sequence is running, error –221,
"Settings conflict;list mode is running" occurs.
[1] Applicable for U2042XA/U2044XA/U2049XA only.
Keysight U2040 X-Series Programming Guide
Range of values
Description/Defaul t value
A numeric value defining the exclusion area offset time.
The default value is 0 s.
0.0 s to 0.1 s
Units are resolved to 50 ns.
Sets the exclusion area offset time to 0.01 s.
SENSe Subsystem
[1]
16
257