Table 1-5 Bit Definitions - Questionable Status Registers; Table 1-6 Bit Change Conditions For Questionable Status Register - Keysight Technologies U2040 X-Series Programming Manual

Wide dynamic range power sensors
Table of Contents

Advertisement

Questionable Status register
The questionable status register provides information about the quality of the
U2040 X-Series measurement results. Any or all of these conditions can be reported in the
questionable data summary bit through the enable register. You must write a value using
the STATus:QUEStionable:ENABle command to set the enable register mask.
The following bits in these registers are used by the U2040 X-Series.
Table 1-5
Bit number
0 to 2
3
4 to 7
8
9
10 to 14
15
The condition bits are set and cleared under the following conditions:
Table 1-6
Bit change conditions for questionable status register
Bit number
Meaning
3
POWer Summary
8
CALibration Summary
9
Power-On Self-Test
Keysight U2040 X-Series Programming Guide
Bit definitions - Questionable status registers
Decimal weight
Definition
-
Not used
8
POWer Summary
-
Not used
256
CALibration Summary
512
Power-On Self-Test
-
Not used
-
Not used (always 0)
EVENts causing bit changes
This is a summary bit for the Questionable POWer Register.
– SET:
Error –230, "Data corrupt or stale"
Error –231, "Data questionable;Input Overload"
Error –231, "Data questionable;CALC1 log error"
Error –231, "Data questionable;CALC2 log error"
Error –231, "Data questionable;CALC3 log error"
Error –231, "Data questionable;CALC4 log error"
– CLEARED: When no errors are detected by the U2040 X-Series during a measurement covering
the causes given for it to set.
This is a summary bit for the Questionable CALibration Register.
– SET:
These may be caused by CALibration[1]:ZERO:AUTO ONCE or
CALibration[1]:AUTO ONCE or CALibration[1][:ALL] or
CALibration[1][:ALL]?.
Error –231, "Data questionable;ZERO ERROR"
Error –231, "Data questionable;CAL ERROR"
– CLEARED: When any of the commands listed above succeed and no errors are placed on the
error queue.
– SET: This bit is set when the power-on self-test fails.
– CLEARED: When the power-on self-test passes.
U2040 X-Series Remote Operation
1
61

Advertisement

Table of Contents
loading

Table of Contents