INITiate Subsystem

INITiate[1]:CONTinuous:SEQuence[1]

Syntax
INITiate[1]:CONTinuous:SEQuence[1] ON|OFF|1|0
INITiate[1]:CONTinuous:SEQuence[1]?
Description
This command sets the U2040 X-Series for either a single trigger cycle or continuous
trigger cycles. A trigger cycle means that the U2040 X-Series exits the wait-for-trigger
state and starts a measurement.
If INITiate:CONTinuous:SEQuence[1] is set to:
– OFF, the trigger system remains in the idle state until it is set to ON, or
INITiate:IMMediate is received. Once this trigger cycle is complete, the trigger
system returns to the idle state.
– ON, the trigger system is initiated and exits the idle state. On completion of each trigger
cycle, the trigger system immediately commences another trigger cycle without
entering the idle state.
The query enters a 1 or 0 into the output buffer.
– 1 is returned when there is continuous triggering
– 0 is returned when there is only a single trigger
This command performs the same function as INITiate[1]:CONTinuous.
NOTE
Example
INIT:CONT:SEQ ON
Reset conditions
– On reset (*RST), this command is disabled.
– On preset (SYSTem:PRESet) and U2040 X-Series power-up, this command is enabled.
148
Initiates the trigger.
Keysight U2040 X-Series Programming Guide