Cable Installation

10.Remove the wire cable hold-down on the right side of the chassis to free the gray coax cables
going to the A16 Reference Assembly.
11.Refer to Figure 21. Locate W38 in the Opt B40/BUN Cable Kit. This cable will have part number
8121-1861 and will be labeled "901" on one end and "15" on the other. Connect the end
labeled "901" to A15J901 on the A15 Enhanced Front End Controller (EFEC) Assembly. Route
the cable along the side of the EFEC board to the right side of the instrument and through the
cutout in the chassis between A9 and A12. Connect the other end to A3J15 on the A3 DIF
Assembly.
Figure 21
W36, W37, W38, and W39 Cables
Installation Note N9010-90065
Installation Procedure
31