Technical Specifications - ResMed AirFit N10 User Manual

Nasal mask
Hide thumbs Also See for AirFit N10:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

Technical Specifications

Pressure-flow
The mask contains passive venting to protect against rebreathing.
As a result of manufacturing variations, the vent flow rate may
curve
vary.
Dead space
Physical dead space is the empty volume of the mask to the end of
the swivel. Using the wide cushions it is 104 mL.
information
Therapy
4 to 30 cm H
2
pressure
Resistance
Drop in pressure measured (nominal)
at 50 L/min: 0.4 cm H
at 100 L/min: 1.2 cm H
Environmental
Operating temperature: +5°C to +40°C
Operating humidity: 15% to 95% non-condensing
conditions
Storage and transport temperature: -20°C to +60°C
Storage and transport humidity: up to 95% non-condensing
Sound
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance
with ISO 4871. The A-weighted sound power level of the mask is
31 dBA, with uncertainty of 3 dBA. The A-weighted sound pressure
level of the mask at a distance of 1 m is 23 dBA, with uncertainty of
3 dBA.
483 mm (H) x 155 mm (W) x 110 mm (D)
Gross
Mask fully assembled with short tube assembly (no headgear)
dimensions
Mask setting
For S9: Select 'Pillows'
For other devices: Select 'SWIFT' (if available), otherwise select
options
'MIRAGE' as the mask option.
Pressure
(cm H
O
O
2
O
2
Flow
O)
(L/min)
2
4
20
8
29
11
35
12
37
16
43
17
44
20
49
22.5
53
24
54
30
61
English