HOLD FUNCTION,
The hold function may be set to automatic, manual or off by the user. If the hold
function is disabled the density function will not work. If it is set to auto the balance
will hold the weight display when it becomes stable. If it is set to manual the user
must press the [Hold/Print] key to set the hold function.
AU Hd
Off Hd
On Hd
1) Switch on the balance then press and hold the [Unit/Clr] key during the self-
checking test of the display.
2) After a few seconds, the display will show the first function F1 UNT.
3) Press the [Air/Liquid] key to select F2 Hld .
4) Press the [Tare] key to view the current setting for the hold function.
5) To change the setting press the [Air/Liquid] key to scroll through and select a
new setting as shown above.
6) Press [Tare] to confirm and move to the next setting.
7) When F3 bL is displayed press the [Tare] key to enter the parameter, the
[Unit/Clr] key to return to weighing, or press [Air/Liquid] to go to the next
function.
F2 F2 F2 F2 H H H H
L L L L D D D D
Sets the hold function to operate automatically when a weight is
stable.
Sets the hold function to off. The Density function will not
operate. The [Hold/Print] key will function only as a print key.
Sets the hold function to be on when the user presses the
[Hold/Print] key.