All devices are thoroughly tested and checked for order compliance prior to shipping. Upon receipt of the device, check it for shipping damage. If any problem comes to light, contact your local Spirax Sarco representative. Please describe the problem and indicate type and serial number of the device. We extend no guarantee of any kind for repair work that is undertaken without notifying us in advance of the intention to carry out such work.
Installation, commissioning, operating personnel Mechanical and electrical installation, as well as commissioning, maintenance and operation, are to be realized solely by qualified personnel that are authorized by the installation operator to perform such work. All such personnel must read and understand the content of the applicable operating instructions before working with the device.
2.5 Proper use of the device Warning The operator is responsible for ensuring that the material used in the sensor and housing is suitable and that such material meets the requirements for the fluid being used and the ambient site conditions. The manufacturer accepts no responsibility in regard to such material and housing.
ELM DN50 ELM DN150 – DN200 The ELM flow meter range has been designed for use on water / condensate which are in Group 2 liquids of the above mentioned Pressure Equipment Directive. The product’s use on other liquids in Group 1 or Group 2 may be possible but, if this is contemplated, Spirax Sarco should be contacted to confirm the suitability of the product for the application being considered.
Page 7
Packaging, storaging, transport Be careful not to damage the device while unpacking it. The device should be stored in a clean, dry room until it is installed so as to prevent particulate matter from entering the device. Make certain that the ambient temperature in the room in which the device is stored lies within the prescribed range.
3. General product information 3.1 Manufacturer Manufactured for Spirax Sarco by Heinrichs Messtechnik GmbH Robert-Perthel-Str. 9 · D-50739 Köln Telephone: +49 221 49708 – 0, Fax: +49 221 49708 - 178 Internet: http://www.heinrichs.eu, Email: mailto:info@heinrichs.eu 3.2 Product type Magnetic-inductive flowmeter based on Faraday’s law of induction 3.3 Product name...
Page 9
4. Steps prior to operation It is essential that you read these operating instructions before installing and operating the device. The device is to be installed and serviced by a qualified technician only. No instructions, wiring diagrams, and/or supplied software, or any portion thereof, may be produced, stored, in a retrieval system or transmitted by any means, electronic, mechanical, photocopying or otherwise, without prior written permission.
6.2 System design The electromagnetic ELM-*** flowmeter consists of a sensor, which picks up an induced measuring signal from the medium flowing through the pipe, and a transmitter which transforms this signal into standardized output signals (4-20 mA or pulses). The sensor is installed in the pipe while the transmitter is mounted directly on the sensor.
7. Performance characteristics and environmental conditions 7.1 Measuring accuracy 7.1.1 Measured error +/- [0.3% of actual value + 0.0001 * (Q at 10 m/s)] Accuracy Flow velocity v [m/s] Fig. 4 7.1.2 Repeatability +/- [0.15% of actual value + 0.00005 * (Q at 10 m/s)] 7.1.3 Reference conditions In accordance with DIN EN 29104 Fluid temperature 22°C ±...
7.5 Ambient temperature range -20°Celsius to + 60°Celsius (-4°F to 140°F) In the case of an outdoor installation, the device must be protected against direct solar irradiation with a weather shield. 7.6 Storage temperature - 25°Celsius to + 60°Celsius (-13°F to 140°F) 7.7 Degree of protection SG2b standard housing, IP68 (NEMA 6P).
7.8 Process conditions 7.8.1 Fluid temperature The data sheet/rating plate of the connected transmitter must be observed. With directly mounted transmitter on the sensor the heat entry must be considered from the process to the transmitter. 7.8.2 State of aggregation Liquid 7.8.3 Viscosity No restrictions.
8. Installation/conditions for use 8.1 Receipt of goods and transport 8.1.1 Receipt of goods Check the packaging and contents for damage. Inspect the supplied goods to ensure complete delivery and compare the consignment with your order specifications. 8.1.2 Transport The protection caps should only be removed immediately before installation of the device in the pipe.
8.2 Installation conditions The installation location in the pipe must be selected so that the sensor is always fully filled with the fluid and cannot run empty. This can best be guaranteed if it is installed in an ascending pipe or drain. The measuring principle is generally independent of the flow profile of the fluid provided no standing vortices reach into the area where the value is measured, such as downstream from elbows or half-open sliding valves upstream from the sensor.
8.3 Installation Screws, bolts, nuts and seals are not supplied and must therefore be provided by the operator. Install the sensor between the pipes. Please observe the required torques stated Section 8.3.4. The installation of additional grounding rings is described in Section 8.3.3. Use for the flanges only seals in accordance with DIN 2690.
Installation examples Horizontal lining In order to avoid measuring errors caused Installation in a slightly ascending pipe. by gas pockets and lining damage caused by negative pressure, the following points must be observed: Highest point in pipe system. Air bubbles will accumulate in the tube.
Long pipes Always install regulating and shut-off devices downstream from the sensor. (Danger of vacuum!) Fig. 11 Installation of pumps Do not install flowmeters on the suction side of pumps in order to avoid a negative pressure and damage to the tube lining. Fig.
Page 19
8.3.3.1 Grounding the ELM flowmeter Ground Wire: Min 4mm Earth bonding - Uninsulated metal pipes only (not Protective required for plastic pipe installations) Earth Seal Seal Pipe Pipe Pipe Flanges Fig. 13 IM-P309-02 MI Issue 2...
Page 20
8.3.4 Torques for studs and nuts Electromagnetic flowmeters must be installed in the pipe system with special care due to the fact that the flow pipe lining is made of PTFE which is malleable under pressure. If the flange nuts are tightened too much, the sealing surface will deform. If the seals are supposed to function properly, the correct torque is highly important.
Page 21
8.4 Wiring Caution! Installation and wiring may only be performed when the auxiliary power is switched off. Non-compliance can result in electric shock and irreparable damage to electronic parts. 8.5 Nominal size and ranges Volume flow depends on the flow velocity and the nominal size of the flowmeter. The electromagnetic flowmeter has been designed in such a way that it operates within the range of the flow velocities occurring in practical applications.
8.6 Ambient conditions Ambient temperature range For fluid temperatures > 60°C As the sensors are an element of the pipe, these are normally thermally isolated when installed to save energy and prevent accidental physical contact. Due to the process temperature heat is introduced through the support for securing the integral mount transmitter or the terminal box.
8.6.4 Ingress protection The sensor meets the requirements of the protection class IP 67. The following must be observed to ensure compliance with protection class IP 67 when the device has been installed or serviced: The housing seals must be clean and undamaged when placed in the sealing groove. If necessary the seals must be cleaned or replaced.
8.7 Process pressure The maximum permissible process pressure PS is stated on the rating plate and depends on the fluid temperature. PN40 PN25 PN16 PN10 °C Fig. 16 8.8 Fluid temperature The maximum permissible fluid temperature of the device depends on the version and the lining material of the flow tube and is stated on the rating plate.
8.9 Electrical connection Mains: 24 Vdc ±15% Power input: 10 VA 5x20mm DIN 41571-3 Voltage = 24 Vdc Current = 1 AT Fuse: Rated voltage = 250 Vac Breaking capacity = 80A / 250Vac e. g. Fa. Wickmann series 201 8.10 Process terminals Terminals located at the rear side of the transmitter’s housing.
8.11 Output Signal 8.11.1 Isolation All signal outputs are electronically isolated from each other and ground (PE) 8.11.2 Analog output A single 0/4-20mA current output is fitted with optional HART communications and can be configured to output flow rate in volumetric or velocity units. 8.11.3 Pulse output A single pulse output is fitted.
Page 27
8.12 Mains and signal terminals ELM mains and signal terminals Fuse Fig. 18 Process terminals Terminal Label Polarity Function Protective conductor Mains Mains Pulse Pulse output (passive) Pulse Pulse output (passive) Status Status output (passive) Status Status output (passive) Current Out.
8.13 HART connection ® A number of options are available for HART communication. However, for all these options ® loop resistance must be less than the maximum load specified in Section Error! Reference source not found. Error! Reference source not found. (Error! Bookmark not defined. ). The HART -Interface is connected via terminals 8 and 9 of the active current output.
9. Maintenance and repair The transmitter ELM is designed as maintenance-free performance. It contains no parts, which have to be replaced or adjusted cyclically. While commissioning or maintenance, mains power must be switched off. Do not connect or disconnect the wirings between sensor and transmitter while power is on! 9.1 Mains fuse...
Only original replacement parts are to be used. Spirax Sarco accepts no liability for any loss or damage of any kind arising from improper operation of any product, improper handling or use of any replacement part, or from external electrical or mechanical effects, overvoltage or lightning.
11.3 Cathodic protective units Using a cathodic protective unit to avoid corrosion, which put a voltage to the tube wall, it must be connected to terminal FE. The transmitter boards, control panel and internal switches are on the same potential as FE. Warning According to EN 50178:1997 all electrical circuits with protective safety isolation without any protection against contacts must observe the...
12.1 Standard version As standard ELM has 1 active 0/4 … 20 mA analogue output, 1 passive pulse or frequency output and 1 passive status output. 12.2 Display and Keypad A LCD display with backlight is a standard feature.
After the measuring system has been started, the measuring point will continue working with the characteristic values stored in the DSM. Thus, the DSM offers maximum safety and high comfort when exchanging device components. Electronic, Power supply board ELM Slot DSM Fig. 22 At any exchange watch the polarity of the memory chip.
13. Configuration and operation 13.1 Introduction The ELM can be operated depending on equipment by using the HART® interface. In the following, the ELM operation and parameterization using the keypad are described. The keypad is located in the electronic compartment and covered by an inspection window.
1. Display mode: In display mode, measured values can be displayed in various combinations and ELM settings can also be displayed. Parameter settings cannot be changed in this mode. Display mode is the standard (default) operating mode when the device is switched on.
Page 38
Cursor keys: Using the cursor keys, the operator can change numerical values, give YES/ NO answers and select parameters. Each key is assigned a symbol in the following table: Descriptor Symbol Cursor key, arrow to the right > Cursor key, arrow to the left <...
Page 39
Such changes should be kept in a safe place. The ELM customer password in the device when delivered is 0002. The service password allows for modification of all ELM functions. This password is not given to customers.
Page 40
15. ELM functions The software functions of the ELM transmitter are divided into functional classes, are arrayed in a circle and can be navigated by using the < or > cursor keys. To go back to your starting point (the measured values functional class) press Esc.
Page 41
In the following, all software functions that can be accessed using the customer password are described. Functions that are only accessible to the vendor (service functions) are not described in the present document. 15.1 Measured values functional class The measured values functional class contains all functions for displaying the measured values. Fig.
In order to calculate the flow area of the meter tube, enter the inside diameter of the meter tube. To do so, use the "inside diameter" function in the functional class SETTINGS SENSOR + ELM. IM-P309-02 MI Issue 2...
Page 43
15.1.6 Relative flow rate The relative flow rate is the percentage ratio of the (current) volume flow and the entered upper range value of the volume flow. You set this upper range value in the functional class FLOW using the function "volume flow QV URV." The calculation of the relative flow rate is based on the following formula: relative flow rate = 100% x (Qabs –...
15.1.10 Display mode during startup By choosing the Display mode during startup function the operator can define the default display. After the operator switched the device on and did not touch any keys for a longer period of time, the defined default display will be shown. Display mode (QV) According to the description in Section 14.1.3.1 "Selection window / make a selection", one...
15.2 Password functional class The password functional class is comprised of the functions for entering and changing the customer password and entering the service password. To cancel the current action, press Esc. Fig. 28 15.2.1 Customer-password After selecting the Customer password function and pressing enter, the following will be displayed: Password? 0000...
The customer password in the device when delivered is 0002. A valid customer password allows all software parameter changes to be made that are permissible for customers. After the operator switched the device off or did not touch any keys for about 15 minutes, the authorization to change settings related to password entry will automatically be canceled.
Page 47
15.3 Counter functional class The counters functional class is comprised of the following functions: Fig. 29 To change the current settings, enter the customer password. Otherwise, the settings can only be displayed but not changed. To cancel the current action, press Esc. IM-P309-02 MI Issue 2...
15.3.2 Reset counter The ELM has 3 independent totalizing counters. Counter 1 and Counter 2 for forward flow and a reverse flow counter. Each of them can be reset individually on the initial value 0.00.
Page 49
15.4 Measurement processing functional class The measurement processing functional class is comprised of all functions that affect the processing of the measured values. To change the current settings, enter the customer password. Otherwise, the settings can only be displayed but not changed. To cancel the current action, press Esc. Fig.
15.4.1 Damping The damping value is intended to dampen abrupt flow rate changes or disturbances. It affects the measured value display and the current and pulse outputs. It can be set in intervals of 1 second from 1 to 60 seconds. After choosing the Damping value function and pressing enter, the following selection field will be display Damping 03 s...
Page 51
15.4.4 Zero point calibration Using the Zero point calibration function the operator can recalibrate the zero point of your meter in the measuring system. Zero point calibration is to be realized after any installation procedure or after any type of work has been performed on in the pipes near the sensor. Refer also Section 11.4 Zero point calibration.
Page 52
15.5 Flow functional class The FLOW functional class is comprised of functions that affect lower- and upper-range values and the processing of the measured flow rates. In Programming mode (see section 14 Operating modes), i.e. after a password has been entered (see 14.1.3.3 Passwords, 15.2 PASSWORD functional class), the operator can change the settings regarding flow.
Page 53
15.5.1 Volume flow QV unit Using this function, the operator can define the physical unit for all display functions, limit values and the upper-range value of volume flow. After choosing the Volume flow QV unit function and pressing enter, the following selection field will be displa Volume flow QV in (l/h) According to the description in Section 14.1.3.1 Selection window / make a selection, one of...
Page 54
15.5.4 Volume flow limit MIN The MIN limiting value for volume flow can be evaluated via the status output. You enter the value as a percentage of the set upper-range value. If the volume flow is lower than that limit value, the status output will be set in case the corresponding assignment has been made.
Page 55
15.5.7 Density If a mass unit in kg or t is used as flow unit (2.5.1 Volume flow QV unit), the density of the medium must be entered in the unit of g/l. Using the entered density value, the mass flow is calculated from the volume flow measurement.
Page 56
15.6 Pulse output functional class The pulse output functional class is comprised of the functions regarding the pulse output. Fig. 32 IM-P309-02 MI Issue 2...
15.6.1 Pulse or frequency output The Pulse or frequency output function allows the operator to define whether pulses per represent a unit of flow or a frequency between 0 and 1 kHz that represents an analog output over the measuring range. After selecting the frequency setting, the maximum frequency of 1 kHz will be generated when the upper-range value for mass or volume flow is reached (depending on the selected pulse unit).
15.6.4 Pulse width This function allows the operator to change the width of the output pulse to be output. If the pulse width is too large for the actual pulse number, it will be reduced automatically. In this case the warning "Pulse output saturated" will be displayed. After selecting the Pulse width function, press ...
Page 59
15.7.1 Status output active state The status output can be compared to an electrical relay that can function as make or break contact. For safety-relevant applications, the operator will choose the break contact setting so that a power failure or failure of the electronics can be detected like an alarm. In standard applications, the output is used as make contact.
Page 60
15.8 Current output functional class The current output functional class allows the operator to perform the settings for the current outputs of the transmitter. Fig. 34 The current output is always assigned to volume flow. IM-P309-02 MI Issue 2...
Page 61
15.8.1 Current output 0/4 - 20 mA The Current output 0/4 to 20 mA function allows the operator to define the range in which the current output is to be operated. Within the range from 0 to 21.6 mA (= 0 ... 110%) HART®...
Page 62
15.9 Simulation functional class The functional class simulation is comprised of the functions for simulating the outputs. If simulation is activated, all output signals will be generated based on the selected type of simulation. The peripherals connected to the device can be tested without a flowing product. Simulation will be deactivated automatically if the operator switched the device off or did not touch any control unit keys for about 10 minutes.
Page 63
15.9.2 Simulation direct / preset value Q This function allows the operator to define whether simulation is comprised of the measurement of the volume flow or whether the outputs will be set directly. Press to display the selected type of simulation. Simulation [direct] As mentioned in Section 14.1.3.1 Selection window / make a selection, the operator can...
Page 64
15.9.5 Status output simulation The Status output simulation function allows the operator to purposefully activate the status output. Press to display the current state. Status output (off) As mentioned in Section 14.1.3.1 Selection window / make a selection, the operator can toggle between "on"...
Page 65
15.10 Self-test functional class The self-test function class is comprised of the functions relating to the self-test of the sensor. The diagnostic functions of the transmitter, which monitor the proper functioning of the electronics and the software, are always active and cannot be switched off. The excitation current can be monitored in addition.
Page 66
15.10.1 Self-test test on / off The Self-test on/off function allows the operator to activate or deactivate the monitoring function of the field coil current. Self-test (off) According to the description in Section 14.1.3.1 Selection window / make a selection, the operator can toggle between "on"...
Page 67
15.10.3 Reference calibration period (GAP) The function Reference calibration period is a multiplication of the function "self-test period" .With the help of this function, you define after how many STP’s the reference calibration is to be performed. Reference calibration GAP = 540 * STP This field shows the current reference calibration period.
Page 68
15.11 Settings sensor + UMF2 (B) functional class This functional class is comprised of the general settings affecting the behavior of the transmitter. SETTINGS SENSOR + UMF2 functional class SETTINGS SETTINGS SENSOR + UMF2 SENSOR + UMF2 Flow direction [forward & reverse] Sensor constant /mV Flow _________________...
The inside diameter of the sensor connected to the transmitter is necessary for calculating the mean flow velocity. The inside diameter must be checked in the ELM (on mm exact) to ensure a correct measurement. After choosing the function "inside diameter" and pressing...
15.11.4 Language Two languages are available in the ELM: German and English.. Language (English) As mentioned in Section 14.1.3.1 Selection window / make a selection, the operator can toggle between these languages: German English. 15.11.5 Excitation frequency With the help of the function Excitation frequency, you can set the excitation frequency of the field coil current.
-20% -40% -60% -80% -100% 100% Plus% Minus% Measured value Fig. 39 15.11.8 Software version (information field) After selecting this function, the version of the transmitter software will be shown (example: 1.06): Version of ELM 001.06 IM-P309-02 MI Issue 2...
Page 72
The integrated diagnostic system of the ELM distinguishes between two types of errors (see also Section 15.11.12 "ELM error messages"). Self-test errors such as problems with a sensor line or inconsistent parameter inputs are displayed as textual error messages. Once the error has been eliminated, the message automatically disappears from the display.
16. Standard operating mode The transmitter operates as described above. After the cause of the error message has been eliminated, the message automatically disappears. 16.1 List of error messages 16.1.1 Display of self-test errors Self-test errors are displayed as plain text in the set language (German or English) on the second line of the LCD.
Page 74
16.1.1 Display of self-test errors (continued) Display Display Description Possible cause of (German) (English) error and remedy Parameter inkons. params Parameter is Check the parameter inconsist inconsistent. settings. The set parameters are contradictory. Check the parameter settings. The set parameters are contradictory.
16.1.2 Display of system error System errors consist of the message text "system error" and a 5-digit number in hexadecimal code. The meaning of the individual error codes is described in the following table. If several errors occur at the same time, the hexadecimal sum of the individual errors will be displayed. The errors are coded in such a way that the individual errors can be easily identified.
18. Maintenance The device requires no maintenance if used according to its intended purpose. Cleaning might be necessary due to deposits and dirt on the electrodes or the flow tube. IM-P309-02 MI Issue 2...
Name of contact person: ....Phone: ......... Information pertaining to the enclosed flowmeter Model ELM- ....was operated using the following fluid: ........In as much as this fluid is water-hazardous / toxic / corrosive / combustible we have done the following:...
Need help?
Do you have a question about the ELM and is the answer not in the manual?
Questions and answers