Action
Basic version
With increased
vibration resistance
With extended
measuring range
8.5
Measuring accuracy for thermocouples
Tolerance classes
The tolerance classes of the thermocouples are defined in the following table in accordance with IEC 584/DIN EN 60584:
Catalog versions
Type
N
333 °C ... 1100 °C ±0.0075x|t[°C]|
(631 °F ... 2012 °F ±0.0075x|t[°F]-32|)
K
333 °C ... 1000 °C ±0.0075x|t[°C]|
(631 °F ... 1832 °F ±0.0075x|t[°F]-32|)
J
333 °C ... 750 °C ±0.0075x|t[°C]|
(631 °F ... 1382 °F ±0.0075x|t[°F]-32|)
Further base thermocouples
Type
T
133 °C ... 350 °C ±0.0075x|t[°C]|
(271 °F ... 662 °F ±0.0075x|t[°F]-32|)
E
333 °C ... 900 °C ±0.0075x|t[°C]|
(631 °F ... 1652 °F ±0.0075x|t[°F]-32|)
Further noble thermocouples
Type
R,S
600 °C ... 1600 °C ±0.0025x|t[°C]|
(1112 °F ... 2912 °F ±0.0025x|t[°F]-32|)
B
600 °C ... 1700 °C ±0.0025x|t[°C]|
(1112 °F ... 3092 °F ±0.0025x|t[°F]-32|)
SITRANS TSinsert/TS100/TS200/TS300/TS500
A5E32897098-01, 09/2013
Tolerance
Class B
Class A
Class AA (1/3 B)
Class B
Class A
Class AA (1/3 B)
Class B
Class A
Basic accuracy, Class 2
-40 °C ... +333 °C ±2.5 °C
(-40 °F ... +631 °F ±4.5 °F)
-40 °C ... +333 °C ±2.5°C
(-40 °F... +631 °F ±4.5 °F)
-40 °C ... +333 °C ±2.5 °C
(-40 °F ... +631 °F ±4.5 °F)
Basic accuracy, Class 2
-40 °C ... +133 °C ±1 °C
(-40 °F ... +271 °F ±1.8 °F)
-40 °C ... +333 °C ±2.5°C
(-40 °F... +631 °F ±4.5 °F)
Basic accuracy, Class 2
0 °C ... 600 °C ±1.5 °C
(32 °F ... +1112 °F ±2.7 °F)
Precision
Basic accuracy
Increased accuracy
High accuracy
Basic accuracy
Increased accuracy
High accuracy
Basic accuracy
Increased accuracy
375 °C ... 1000 °C ±0.004x|t[°C]|
(707 °F ... 1832 °F ±0.004x|t[°F]-32|)
375 °C ... 1000 °C ±0.004x|t[°C]|
(707 °F ... 1832 °F ±0.004x|t[°F]-32|)
375 °C ... 750 °C ±0.004x|t[°C]|
(707 °F ... 1382 °F ±0.004x|t[°F]-32|)
125 °C ... 350 °C ±0.004x|t[°C]|
(257 °F ... 662 °F ±0.004x|t[°F]-32|)
375 °C ... 800 °C ±0.004x|t[°C]|
(707 °F ... 1472 °F ±0.004x|t[°F]-32|)
1100 °C ... 1600 °C ±[1 + 0.003 x(t -1100)] °C
(2012 °F ... 2912 °F ±1,8+0,003x(t[°F]-2012)
Range [°C (°F)]
-50 ...400 (-58 ... +752)
-30°... 300 (-58 ... +572)
0°... 150 (32 ... 302)
-50°... 400 (-58 ... +752)
-30°... 300 (-58 ... +662)
0°... 150 (32 ... 302)
-196 ... 600 (392 ... 1112)
-196 ... 600 (392 ... 1112)
Increased accuracy, Class 1
-40 °C ... +375 °C ±1.5 °C
(-40 °F ... +707 °F ±2.7 °F)
-40 °C ... +375 °C ±1.5 °C
(-40 °F... +707 °F ±2.7 °F)
-40 °C ... +375 °C ±1.5 °C
(-40 °F ... +707 °F ±2.7 °F)
Increased accuracy, Class 1
-40 °C ... +125 °C ±0.5 °C
(-40 °F ... +257 °F ±0.9 °F)
-40 °C ... +375 °C ±1.5 °C
(-40 °F... +707 °F ±2.7 °F)
Increased accuracy, Class 1
0 °C ... 1100 °C ±1 °C
(32 °F ... +2012 °F ±1.8 °F)
-