Plus Or Minus Cylinder Convention; Diopter Power Resolution - Welch Allyn SureSight Manual

Autorefractor
Hide thumbs Also See for SureSight:
Table of Contents

Advertisement

Plus or Minus Cylinder Convention

Your SureSight was set at the factory to the Minus Cylinder Convention. If you
wish to change your unit to the plus convention:
1) Open the battery cover located on the underside of the unit by sliding it in the
direction indicated by the arrow and lift up.
2) Rotate the lever holding the battery in and lift the battery from its
compartment.
3) View the DIP switches inside the battery compartment located on the button
side of the unit.
4) Using tweezers or another appropriate tool, pull the first switch (located on
the side of the unit with the handstrap) up towards the bottom of the
SureSight.

Diopter Power Resolution

Your SureSight was set at the factory to display power in 1/4 diopter increments.
If you wish to change to 1/8 diopter increments:
1) Open the battery cover located on the underside of the unit by sliding it in the
direction indicated by the arrow and lift up.
2) Rotate the lever holding the battery in and lift the battery from its
compartment.
3) View the DIP switches inside the battery compartment located on the button
side of the unit.
4) Using tweezers or another appropriate tool, pull the third switch (third from
side of SureSight with the handstrap) up towards the bottom of the
SureSight.
14

Advertisement

Table of Contents
loading

Table of Contents