TROUBLESHOOTING

Power on
but no display
Unstable
readings
Slow response
time or reading
drift
Dry bulb
Static charge
Same readings
in different
buffers and
samples
Batteries
Batteries
Batteries
Electrode
Electrode
Electrode
Electrode
Electrode
Clogged junction
Strong alkaline
measurement
Deteriorated gel layer
Protein coating on
electrode surface
Oil, paint, dyes,
suspended solids on
electrode
Dehydrated membrane,
long term storage
without wetting
Wiping electrode
Cracked or broken
electrode
Insert batteries
Verify correct polarity
Replace batteries
Immerse electrode more deeply
in sample
Condition electrode before first
use
Remove air bubbles caught
under electrode
Clean electrode
Replace electrode
Soak in 4.07M Potassium
Chloride (KCl) at 60°C for 30
minutes
Soak in 0.1M Hydrochloric acid
(HCl) overnight
Replace electrode
Soak in 1g Pepsin dissolved in 10
mL of 0.1M HCl for 30 minutes or
as needed
Rinse electrode alternately with
materials solvent then a buffer
7.00
Soak electrode tip in wetting cap
filled with 1 mL 4.0 buffer for
24-48 hours
Rinse electrode in 4.0 buffer and
blot. Do not wipe electrode
Replace electrode. Use bulb
guard. Avoid plunging electrode
into bottom of container and stir
bars.