LaMotte 1200-MN User Manual

1200 colorimeter manganese

Advertisement

Quick Links

QUANTITY
2 x 100 mL
2 x 30 mL
15 mL
1
1
6
1
1
*WARNING: Reagents marked with an * are considered hazardous substances. To view or print a
Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or our web site. To obtain a
printed copy, contact us by e-mail, phone, or fax.
To order individual reagents or test kit components, use the specified code number.
INTRODUCTION
Manganese is present in ground water in the divalent state due to the lack of oxygen. In
surface waters, manganese may be in various oxidation states as soluble complexes or as
suspended compounds. Manganese is rarely present in excess of 1 mg/L. It may cause an
objectionable taste or cause staining problems in laundry, but manganese levels normally
encountered in water seldom produce any health hazard.
Manganese is removed from water by various means including chemical precipitation,
pH adjustment, aeration, superchlorination and the use of ion exchange resins.
1200 COLORIMETER
MANGANESE
MODEL 1200-MN · CODE 3682-01
CONTENTS
*Hardness Buffer Reagent
*Manganese Indicator Reagent
*Sodium Cyanide, 10%
Pipet, 0.5 mL, plastic, w/cap
Pipet, 1.0 mL, plastic
Colorimeter Tubes, with caps
Water Sample Collecting Bottle
1200 Colorimeter for Manganese
CODE
*4255-J
*3956-G
*6565-E
0369
0354
0290-6
0688
26739

Advertisement

loading
Need help?

Need help?

Do you have a question about the 1200-MN and is the answer not in the manual?

Questions and answers

Summary of Contents for LaMotte 1200-MN

  • Page 1 1200 COLORIMETER MANGANESE MODEL 1200-MN · CODE 3682-01 QUANTITY CONTENTS CODE 2 x 100 mL *Hardness Buffer Reagent *4255-J 2 x 30 mL *Manganese Indicator Reagent *3956-G 15 mL *Sodium Cyanide, 10% *6565-E Pipet, 0.5 mL, plastic, w/cap 0369 Pipet, 1.0 mL, plastic...
  • Page 2 MANGANESE TEST PROCEDURE - PAN METHOD Read the 1200 Colorimeter Manual before proceeding. Carefully wipe tubes dry before inserting into the colorimeter chamber. MANGANESE Rinse and fill a colorimeter tube Fill the Water (0290) to the Sample Collecting 10 mL line with Bottle (0688) with sample water.
  • Page 3 Use the 0.5 mL pipet (0369) to add 0.5 mL of Cap tube and invert *Manganese to mix. Indicator Reagent (3956). Align the index line with the Wipe tube dry. arrow on the meter, insert L Mott tube into chamber. Close the lid.
  • Page 4 None. Test is quite specific. LaMOTTE COMPANY Helping People Solve Analytical Challenges ® PO Box 329 • Chestertown • Maryland • 21620 • USA 800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax: 410-778-6394 Visit us on the web at www.lamotte.com 63682-01 · 10.04...

This manual is also suitable for:

3682-01