Section 6

6.1 Atmospheric pressure equilibrium

Since the instrument is sealed against moisture, you must open the barometric pressure sensor
relief valve switch on top of the instrument (refer to
permit the instrument to achieve atmospheric pressure equilibrium, and take an accurate
barometric pressure reading.
This must be done with every calibration. To open the relief valve, depress and hold down the
pressure sensor relief valve switch for five seconds, then release.

6.2 Pressure calibration

If you have access to an accurate barometer, you may wish to calibrate the instrument's internal
barometric pressure sensor. This is done using the PC WinLog97 program. Choose
Troubleshooting, Pressure Calibration and an informational message will appear to warn you
that the instrument's current pressure calibration will be lost.
Choose OK to continue. The calibration procedure then displays a Pressure Calibration dialog
box. The Measured Pressure value shows the current instrument pressure reading.
Note: Since the instrument is sealed against moisture, you must open the barometric pressure sensor relief
valve switch (location depicted in
pressure equilibrium, and take an accurate barometric pressure reading.
Enter the current atmospheric pressure, in mbars, in the Calibration Pressure entry box.
Choose Calibrate to direct the instrument to read and display the Measured Pressure using
this calibration value.
Choose Quit when you are satisfied with the pressure calibration to return to normal operation.

6.3 Calibration range checking

When calibrating the sensor in air or in a span gas, as detailed in
new calibration current should be between 25% to 175% of the ideal current (which is stored in
the instrument's non-volatile memory for each membrane). If not, the instrument displays Err on
its LCD and the system will not calibrate. When the system will not calibrate, it is likely that a
sensor service will be required.
Refer to
Calibrations
Troubleshooting on page
Figure 1 on page 17
Figure 1 on page
17) to permit the instrument to achieve atmospheric
43, if you suspect any problems with the analyzer.
for actual location) to
Sensor calibration
below, the
39