Measuring Accuracy For Thermocouples - Siemens SITRANS TS100 Compact Operating Instructions

Temperature sensors
Table of Contents

Advertisement

8.5

Measuring accuracy for thermocouples

Tolerance classes
The tolerance classes of the thermocouples are defined in the following table in accordance
with IEC 584/DIN EN 60584:
Catalog versions
Type
N
333 °C ... 1100 °C ±0.0075x|t[°C]|
(631 °F ... 2012 °F ±0.0075x|t[°F]-32|)
K
333 °C ... 1000 °C ±0.0075x|t[°C]|
(631 °F ... 1832 °F ±0.0075x|t[°F]-32|)
J
333 °C ... 750 °C ±0.0075x|t[°C]|
(631 °F ... 1382 °F ±0.0075x|t[°F]-32|)
Further base thermocouples
Type
T
133 °C ... 350 °C ±0.0075x|t[°C]|
(271 °F ... 662 °F ±0.0075x|t[°F]-32|)
E
333 °C ... 900 °C ±0.0075x|t[°C]|
(631 °F ... 1652 °F ±0.0075x|t[°F]-32|)
Further noble thermocouples
Type
R,S
600 °C ... 1600 °C ±0.0025x|t[°C]|
(1112 °F ... 2912 °F ±0.0025x|t[°F]-32|)
B
600 °C ... 1700 °C ±0.0025x|t[°C]|
(1112 °F ... 3092 °F ±0.0025x|t[°F]-32|)
TS100/TS200/TS300/TS500/TSinsert/TSthermowell
Compact Operating Instructions, 01/2018, A5E03920348-AE
Basic accuracy, Class 2
-40 °C ... +333 °C ±2.5 °C
(-40 °F ... +631 °F ±4.5 °F)
-40 °C ... +333 °C ±2.5°C
(-40 °F... +631 °F ±4.5 °F)
-40 °C ... +333 °C ±2.5 °C
(-40 °F ... +631 °F ±4.5 °F)
Basic accuracy, Class 2
-40 °C ... +133 °C ±1 °C
(-40 °F ... +271 °F ±1.8 °F)
-40 °C ... +333 °C ±2.5°C
(-40 °F... +631 °F ±4.5 °F)
Basic accuracy, Class 2
0 °C ... 600 °C ±1.5 °C
(32 °F ... +1112 °F ±2.7 °F)
8.5 Measuring accuracy for thermocouples
Increased accuracy, Class 1
-40 °C ... +375 °C ±1.5 °C
(-40 °F ... +707 °F ±2.7 °F)
375 °C ... 1000 °C ±0.004x|t[°C]|
(707 °F ... 1832 °F ±0.004x|t[°F]-32|)
-40 °C ... +375 °C ±1.5 °C
(-40 °F... +707 °F ±2.7 °F)
375 °C ... 1000 °C ±0.004x|t[°C]|
(707 °F ... 1832 °F ±0.004x|t[°F]-32|)
-40 °C ... +375 °C ±1.5 °C
(-40 °F ... +707 °F ±2.7 °F)
375 °C ... 750 °C ±0.004x|t[°C]|
(707 °F ... 1382 °F ±0.004x|t[°F]-32|)
Increased accuracy, Class 1
-40 °C ... +125 °C ±0.5 °C
(-40 °F ... +257 °F ±0.9 °F)
125 °C ... 350 °C ±0.004x|t[°C]|
(257 °F ... 662 °F ±0.004x|t[°F]-32|)
-40 °C ... +375 °C ±1.5 °C
(-40 °F... +707 °F ±2.7 °F)
375 °C ... 800 °C ±0.004x|t[°C]|
(707 °F ... 1472 °F ±0.004x|t[°F]-32|)
Increased accuracy, Class 1
0 °C ... 1100 °C ±1 °C
(32 °F ... +2012 °F ±1.8 °F)
1100 °C ... 1600 °C ±[1 + 0.003 x(t -1100)] °C
(2012 °F ... 2912 °F ±1,8+0,003x(t[°F]-2012)
-
Technical data
59

Advertisement

Table of Contents
loading

Table of Contents