8.4 Measuring tolerances for resistance thermometers

8.4
Measuring tolerances for resistance thermometers
Tolerance classes
The tolerance classes of the resistance thermometers are defined as follows in accordance
with IEC 60751:
Tolerance class
Class B
Class A
Class AA (1/3 B)
Tolerances
The following tables provide an overview of the validity ranges of these tolerances. When
you use a thermometer above the specified limits, the values of the next lower accuracy
class apply.
Action
Basic version
With increased vibra-
tion resistance
With extended measur-
ing range
If an RTD will be subjected in a higher temperature as specified in the class, the sensor will
be declassified to the referring accuracy class. For enduring operation near 400 °C we
recommend the use of an extended measuring range RTD, instead of basic and increased
vibration versions.
58
Basic accuracy
Increased accuracy
High accuracy
Tolerance
Class B
Class A
Class AA (1/3 B)
Class B
Class A
Class AA (1/3 B)
Class B
Class A
Class AA (1/3 B)
Precision
Precision
Basic accuracy
Increased accuracy
High accuracy
Basic accuracy
Increased accuracy
High accuracy
Basic accuracy
Increased accuracy
Increased accuracy
TS100/TS200/TS300/TS500/TSinsert/TSthermowell
Compact Operating Instructions, 01/2018, A5E03920348-AE
∆t
±(0.30 °C +0.0050|t[°C]|)
±1.8x0.30 °F +0.0050x|t[°F]-32|
±(0.15 °C +0.0020|t[°C]|)
±1.8x0.15 °F +0.0020x|t[°F]-32|
±(0.10 °C +0.0017|t[°C]|)
±1.8x0.10 °F +0.0017x|t[°F]-32|
Range
-50 ... +400 °C (-58 ... +752 °F)
-30 ... +300 °C (-22 ... +572 °F)
0 ... +150 °C (+32 ... +302 °F)
-50 ... +400 °C (-58 ... +752 °F)
-30 ... +300 °C (-22 ... +572 °F)
0 ... +150 °C (+32 ... +302 °F)
-196 ... +600 °C (-320 ... +1112 °F)
-100 ... +450 °C (-148 ... +842 °F)
-50 ... +250 °C (-58 ... +482 °F)