Figure 3-3. Vxibus Slot Configuration - National Instruments MXI VXI-PCI8026 Getting Started

Ni-vxi/ni-visa software for linux
Table of Contents

Advertisement

Chapter 3
VXI-MXI-2 Configuration and Installation
VXI/VME-PCI8026 for Linux
a. Automatic Slot 0 Detection (Default)
c. Manual Nonslot 0 Configuration

Figure 3-3. VXIbus Slot Configuration

When the VXI-MXI-2 is installed in Slot 0, it becomes the VMEbus
System Controller. In this role, it has VMEbus Data Transfer Bus Arbiter
circuitry that accepts bus requests on all four VMEbus request levels,
prioritizes the requests, and grants the bus to the highest priority requester.
As VMEbus System Controller, the VXI-MXI-2 also drives the 16 MHz
VMEbus system clock by an onboard 16 MHz oscillator.
As required by the VXIbus specification, the VXI-MXI-2 drives the
10 MHz signal CLK10 on a differential ECL output when installed in
Slot 0. When not installed in Slot 0, the VXI-MXI-2 only receives the
CLK10 signal.
W2
Slot 0
Auto
Nonslot 0
W2
Slot 0
Auto
Nonslot 0
b. Manual Slot 0 Configuration
W2
Slot 0
Auto
Nonslot 0
3-6
www.ni.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mxi vme-pci8026

Table of Contents