4
Calibration Procedures

0 dB Range Flatness Verification

This procedure checks the high frequency ac amplitude flatness above 100 kHz on
the 0dB attenuator range. (Flatness is relative to 1 kHz.)
1 Connect the power meter to measure the output amplitude of the instrument
2 Set up the function generator as follows:
3 On the power meter, use the Relative Power function to set the current reading
4 Set the power meter offset to equal the 100kHz_0dB_offset value previously
72
as shown below.
– Output impedance: 50 Ω (press
– Waveform: Sine
– Frequency: 100 kHz
– Amplitude: 3.51 dBm
Make sure the output is enabled.
as the reference value. This will allow you to compare future measurement
results in dB.
calculated. This sets the power meter to directly read the flatness error
specification relative to 1 kHz. 100kHz_0dB_offset is calculated on the
"Amplitude and Flatness Verification
and select Output Setup).
Worksheet".
Keysight 33210A Service Guide