0 dB Range Flatness Adjustments

1 Connect the power meter as shown on page 98.
2 Use the power meter to measure the output amplitude for each of the setups
Setup 39 establishes the power meter reference for all the remaining setups in
NOTE
this table. You must always perform setup 39 before any of the following setups.
96
in the table below.
Nominal Signal
Setup
Frequency
[a]
100 kHz
39
[a]
200 kHz
40
[a]
500 kHz
41
[a]
1.5 MHz
42
[a]
3 MHz
43
[a]
4 MHz
44
[a]
6 MHz
45
[a]
8 MHz
46
[a]
9.9 MHz
47
[a]
200 kHz
48
[a]
500 kHz
49
[a]
1.5 MHz
50
[a]
3.0 MHz
51
52
53
[a] Constants are stored after completing this setup.
Amplitude
0.28 Vrms
2 dBm
Power Meter Reference for 0dB Range
0.28 Vrms
2 dBm
Flatness for 0 dB, Elliptical Filter
0.28 Vrms
2 dBm
Flatness for 0 dB, Elliptical Filter
0.28 Vrms
2 dBm
Flatness for 0 dB, Elliptical Filter
0.28 Vrms
2 dBm
Flatness for 0 dB, Elliptical Filter
0.28 Vrms
2 dBm
Flatness for 0 dB, Elliptical Filter
0.28 Vrms
2 dBm
Flatness for 0 dB, Elliptical Filter
0.28 Vrms
2 dBm
Flatness for 0 dB, Elliptical Filter
0.28 Vrms
2 dBm
Flatness for 0 dB, Elliptical Filter
0.28 Vrms
2 dBm
Flatness for 0 dB, Linear Phase Filter
0.28 Vrms
2 dBm
Flatness for 0 dB, Linear Phase Filter
0.28 Vrms
2 dBm
Flatness for 0 dB, Linear Phase Filter
0.28 Vrms
2 dBm
Flatness for 0 dB, Linear Phase Filter
Setup not used for this instrument
Setup not used for this instrument
Keysight 33210A Service Guide