Keysight Technologies 33210A Service Manual page 99

10 mhz function/arbitary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

3 Using the numeric keypad, adjust the displayed amplitude at each setup to
In order to get dBm you must use the numeric keypad (not the knob) to enter the
NOTE
number, and then select "dBm".
4 After performing setup 62:
Keysight 33210A Service Guide
Nominal Signal
Setup
Frequency
[a]
500 kHz
56
[a]
1.5 MHz
57
[a]
3.0 MHz
58
59
4.0 MHz
[a]
6.0 MHz
60
[a]
8.0 MHz
61
[a]
9.9 MHz
62
[a] Constants are stored after completing this setup.
match the measured amplitude (in dBm). Then select ENTER VALUE.
a If your calibration procedures require you to verify the adjustment just
made, exit the calibration menu and perform
Verification"
on page 74.
b If you are making all the adjustments and then verifying the instrument's
performance, continue with the next procedure in this chapter.
Amplitude
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
0.9 Vrms
12 dBm
Flatness for 10 dB, Elliptical Filter
Calibration Procedures
"+10 dB Range Flatness