Siemens SITRANS F Series Function Manual page 37

Profibus (from firmware 4.0) coriolis flowmeters
Hide thumbs Also See for SITRANS F Series:
Table of Contents

Advertisement

Index
Parameter name
81
MASS_FLOW_HI_LI
MIT
82
DENSITY
83
DENSITY_UNITS
84
DENSITY_LO_LIMIT
85
DENSITY_HI_LIMIT
86
TEMPERATURE
87
TEMPERA‐
TURE_UNITS
88
TEMPERA‐
TURE_LO_LIMIT
89
TEMPERA‐
TURE_HI_LIMIT
98
LOW_FLOW_CUT‐
OFF
99
FLOW_DIRECTION
FCT030 Profibus (From firmware 4.0)
Function Manual, 06/2018, A5E39931906-AB
Data type Access
Annotation
Float32
Read
Mass flow high limit
101
Read
Process value
Un‐
Read /
Density unit
signed16
Write
The coding is according the PROFIBUS PA Profile specification:
1097: kg/m³
1098: Mg/m³
1099: kg/dm³
1100: g/cm³
1101: g/m³
1102: t/m³
1103: kg/l
1104: g/ml
1105: g/l
1106: lb/in³
1107: lb/ft³
1108: lb/gal
1109: STon/yd³
1430: lb/i.gal
1558: mg/l
1559: µg/l
1564: mg/dm³
1566: mg/m³
1995: customized unit
Float32
Read
Fixed to 0.0
Float32
Read
Fixed to 5000
101
Read
Media temperature
Un‐
Read /
Temperature unit that is used by temperature values
signed16
Write
Coding:
1000: K
1001: °C
1002: °F
1003: °R
Float32
Read
Media temperature flow low limit
Float32
Read
Media temperature high limit
Float32
Read /
Mass flow low-flow cut-off filter
Write
Un‐
Read /
Assigns an arbitrary positive or negative sign to the measured
signed8
Write
PV value.
Coding:
0: positive
1:
negative
Communication
2.10 Acyclic communication
37