2. Use the Navigation buttons to scroll to "Print" and then press the Select button.
To print a previous reading:
1.
Press the Memory button; the most recent reading is displayed.
2. Use the Navigation buttons to move to "Scroll" and then press the Select button.
3. Use the Navigation buttons to scroll to the appropriate reading and press the Select
button.
4. Use the Navigation buttons to scroll to "Print" and then press the Select button.

Weight Scale

The Healthometer scale is not sold with an RS-232 cable to connect the scale to the Spot
Vital Signs LXi. The customer must purchase a standard RS-232 cable.
If the weight scale is enabled in the Internal Configuration Mode, you cannot manually
enter the patient's weight. The Spot LXi supports the Healthometer
www.healthometer.com), Tanita (http://www.tanita.com/) weight scales.
1.
Follow the weight scale's instructions to zero out the weight scale.
2. Instruct the patient to step on the scale and to remain relatively still until the weight
levels out. The Spot LXi display shows the weight. The patient may step off the scale;
the weight will remain on the display.

Barcode Scanner

The Internal Configuration Mode allows you to enable the scanning of patients' and/or
clinicians' barcodes for identification purposes. The barcode scanner supports most linear
barcodes.
1.
Remove the barcode scanner from its holder.
2. Hold the scanner approximately 6 inches (15.4 cm) from the barcode and squeeze the
button so that the light from the scanner appears on the barcode. The scanner
provides an audible tone with each successful barcode reading.
If the scanner has difficulty reading the barcode, slowly adjust the distance and the
angle between the scanner and the barcode while squeezing the scanner button. If it
continues to have difficulty, verify that the barcode is as flat as possible.
The patient identification number displays in the Navigation Window on the Spot LXi
display. However, if both Patient ID and Clinician ID are enabled in the Internal
Configuration Mode, Spot LXi asks if the scanned ID is for the patient or the clinician.
If Patient ID is enabled and the Clinician ID is disabled in the Internal Configuration
Mode, Spot Vital Signs LXi interprets the scanned bar code as the Patient ID.
Caution Only use a 9V battery with the weight scale. Do not use the weight
scale's external power supply.
Caution When using the Healthometer scale, remove the two hex nuts on the
RS-232 cable, as supplied, before screwing the cable into the scale base.