Error Message
-314,"Save/recall memory lost"
-315,"Configuration memory lost"
-330,"Self-test failed"
-350,"Queue overflow"
-360,"Communication error"
-361,"Parity error in program
message"
-362,"Framing error in program
message"
-363,"Input buffer overrun"
-400,"Query error"
-410,"Query INTERRUPTED"
-420,"Query UNTERMINATED"
-430,"Query DEADLOCKED"
1000,"VXI signal queue overflow"
1001,"VXI interrupt queue overflow"
1002,"VXI bus error"
1003,"VXI bus timeout"
1005 - 1999, "Internal error: xxx"
Error Messages B-4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table B-1, Error Messages (Continued)
ESR Bit Set
DDE
DDE
DDE
DDE
DDE
DDE
DDE
DDE
QYE
QYE
QYE
QYE
DDE
DDE
DDE
DDE
DDE
Error Description
The *SAV or *RCL command has failed due to a corrupted
nonvolatile memory.
Configuration information has been lost due to a corrupted
nonvolatile memory.
The self-test has failed. If this is set at power-up, execute the
*TST? query to determine what has failed.
An attempt was made to add an error message to a full error
queue. Thus, the error queue has overflowed.
A communication error has occurred over the RS-232
interface.
A parity error has been detected in the command over the
RS-232 interface.
A framing error has been detected in the command over the
RS-232 interface.
The RS-232 interface input buffer has been overrun.
A query error has been detected.
The query has been interrupted (see IEEE-488.2, Sec.
6.3.2.3).
see IEEE-488.2, Sec 6.3.2.2 for a description.
The query has reached a deadlocked state. This happens
when the input buffer is full, the output buffer is full, and a
reply to a query cannot be generated.
The signal queue has overflowed. Sending VXI signals is an
alternative method of generating interrupts.
The queue of pending interrupts to send has overflowed. The
EMS maintains a queue of interrupts which should be sent.
An interrupt cannot be sent until a previous interrupt has been
acknowledged.
A VXIbus error has occurred when the EMS attempted to
write data over the VXIbus.
A timeout has occurred while the EMS was waiting for control
of the VXIbus.
An internal firmware error has been detected. Errors of this
type should be reported to Racal Instruments for correction.