*SAV Command

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
nonvolatile memory location from which to recall the
instrument state. That is, both of the following formats are
accepted:
*RCL
*RCL <location>
If <location> is specified, it must be in the range 0 to 9. If
<location> is not specified, it will default to 1.
The *RCL command recalls the following instrument state
attributes
Upper and/or Lower Tolerances For:
Power Supply Voltages
Fan speed
Ambient Temperature
VXI Slot Temperature Rise
VXI IACKn count limits
VXI BERR count limit
Elapsed time
Trigger Subsystem Configuration:
Trigger Delay and Delay Application
TRIG IN/TRIG OUT connections
Front Panel Display Configuration:
Front Panel Display On/Off State
User-Defined Display Text
RS-232 Interface Configuration:
Baud Rate
Parity
Data Bits
Stop Bits
Note that Status Registers, Condition Registers, Enable
Registers, and Event Registers are NOT affected by the
*RCL command.
The *SAV command will store the present instrument state
into nonvolatile memory. The *SAV command may specify
a nonvolatile memory location into which the instrument
state is saved. That is, both of the following formats are
accepted:
Using The Enhanced Monitoring System 4-39