MSA Ultima Instruction Manual Addendum page 19

Gas monitors with x3 technology
Hide thumbs Also See for Ultima:
Table of Contents

Advertisement

For systems that use six-digit addressing, 4XXXXX:
The first digit is an internal system requirement and does not
appear in the communications packet. The base address is
440000, and the first valid address is 440001.
Table 3-4. ModBUS Factory Configuration Data (Read Only)
DESCRIPTION
Device Type
Firmware Version
Relays Option Installed
Reserved for future use
Date of Mfg Year, Sensor 1
Date of Mfg Month, Sensor 1
Date of Mfg Day, Sensor 1
Date of Mfg Year, Sensor 2
Date of Mfg Month, Sensor 2
Date of Mfg Day, Sensor 2
Date of Mfg Year, Sensor 3
Date of Mfg Month, Sensor 3
Date of Mfg Day, Sensor 3
Full Scale Range-
Default Sensor 1
Full Scale Range-
Default Sensor 2
Full Scale Range-
Default Sensor 3
ADDRESS
POSSIBLE VALUES
Base+1
3(Ultima X3 US)
4(Ultima X3 Europe)
Base+2
0..32767 Integer,
divide by 100 for range
00.00 to 99.99
Base+3
0-Relays not installed,
1-Relays installed
Base+4
Base+5
20XX
Base+6
1..12
Base+7
1..31
Base+8
20XX
Base+9
1..12
Base+10
1..31
Base+11
20XX
Base+12
1..12
Base+13
1..31
Base+14
Single Precision Float
Base+16
Single Precision Float
Base+18
Single Precision Float
3-3