5. Configuration
The device can be configured via DIP switches and jumpers manually or with the
corresponding software "Digalox Manager" via USB interface. When configuring via
DIP switches and jumpers the basic settings can be made computer-independent.
The configuration with the software "Digalox Manager" allows full functionality.
Scaling
In modes where a scaling is necessary to represent the correct measurement value
(for example 60 mV DC for shunt or analogue signal) the upper scale value
corresponds to the primary value of the shunt or the maximum display value of the
analogue signal.
Example:
upper scale value = 250 = primary value shunt
upper scale value = 100 = maximum display value of analogue signal
Configuration
For configuration with software set DIP switch 10 to OFF position.
Please note that the measurement mode selected in "Digalox Manager" is displayed
only until the power supply of the DPM72 is interrupted.
In further operation, the device always shows the measurement mode,
which is selected by the jumpers J4-J6 or the DIP switches 11-12
Configuration via DIP switches and jumpers if only one measurement mode is
required (Manual mode)
1. Set DIP switch 10 to ON position.
2. Set measurement mode with jumpers at J4-J6 according to table "Measurement
mode", column "Manual mode".
3. Set upper scale caption for the first measurement mode with DIP switches 1-9
(see "Configuration of the upper scale caption").
4. Connect supply voltage.
via
USB
interface
(see table "Measurement mode")
with
software
respectively!