The sensor must be installed at a largely vibration-free location.
The sensor and the converter must be assigned properly for the model FSM4000: The
sensors have an end number of X1, X2, etc., on the name plate. The transmitters have the
end numbers Y1, Y2, etc. End numbers X1 and Y1 are considered a unit.
Monitoring of pulse output.
The pulse output can be operated as active output (24 VDC pulse) or as passive output
(optocoupler). The current setting is provided on the name plate for the transmitter.
Modification as shown in the following illustration.
Fig. 36: Setting the pulse output using jumpers
1 Display board
Important
The transmitter is immediately ready for operation after mounting and installation of the
connections. The parameters are set at the factory.
The connected wires must be checked for firm seating. Only firmly seated wires ensure full
functionality.
5.1.1

Switching on supply power

After switching on the supply power, the flowmeter sensor data in the external FRAM is
compared with the data saved internally. If the data is not identical, the transmitter data is
replaced automatically. Once completed, Warning 7 "Flowmeter sensor data loaded" and
Warning 8b "Update external FRAM" are displayed. The measuring equipment is now ready for
operation.
The display shows the current flowrate.
Important
For information about menu navigation for the device, refer to the the "Parameterization"
section of the operating instructions.
D184B141U02
1
A
A Pulse passive (jumper inner)
B Pulse active (jumper outer)
FSM4000
Startup Operation
B
G00559
EN - 29