Maintenance; Maintenance Intervals - Drager X-am 5000 Technical Manual

Multi-gas monitor approved as type mqg 0010
Hide thumbs Also See for X-am 5000:
Table of Contents

Advertisement

Maintenance

Maintenance

Maintenance intervals

The device should be inspected and maintained by suitably qualified persons annually.
Comparisons:
EN 60079-29-2 – Gas detectors - Selection, installation, use and maintenance of
detectors for flammable gases and oxygen
EN 45544-4 – Electrical apparatus used for the direct detection and direct
concentration measurement of toxic gases and vapours - Part 4: Guide for selection,
installation, use and maintenance
National regulations
Recommended calibration interval for measuring channels Ex, O
CO: 6 months.
Calibration interval of other gases: refer to the Instructions for Use of the respective
DrägerSensors.
Depending on device configuration:
Replace the alkaline batteries or charge the battery – refer to page 55 to page 56 –
after each use, at the latest after the battery alarm has been triggered or after 2
weeks.
Calibrating the device – page 45.
In regular intervals, according to the sensors used and the operating conditions. For
sensor-specific calibration data, refer to the Instructions for Use/data sheets of the
1)
sensors used
.
Before you carry out safety-related relevant measurements, the zero point and
sensitivity of the devices should be tested in accordance with national regulations.
Inspection by suitably qualified persons – every year.
The inspection intervals must be established in each individual case and shortened
if necessary, depending on technical safety considerations, engineering conditions
and the technical requirements of the equipment.
We recommend that a service agreement be concluded with Dräger and that repairs
also be carried out by them.
Replace the sensors, page 59 – if necessary, when it is not possible to calibrate the
sensors anymore.
1)
Instructions for Use/data sheets for DrägerSensors can be downloaded on the product page of the X-am
5000 at the following Internet address: www.draeger.com. See also the enclosed Instructions for Use and
data sheets for the sensors used.
38
, H
S, SO
, NO
and
2
2
2
2

Hide quick links: