Special symbol »
and displayed numerical
code:
601
602
603
604
605
606
607
608
624
626
«
Cause
The zero point calibration of
DrägerSensor XXS EC4 is not
valid
The span calibration of
DrägerSensor XXS EC4 is not
valid
The measured value of
DrägerSensor XXS EC4 is in
the negative range
DrägerSensor XXS EC4 is not
inserted or defective
Error during the function test
with gas (bump test) of
DrägerSensor XXS EC4
Failed filter test.
Failed rise time test.
User parameters not feasible Check configuration of user
Device incorrectly configured
by Dräger CC-Vision.
Error during warm-up acceler-
ation Dräger Sensor XXS
EC4
Faults, Cause and Remedy
Remedy
Carry out fresh air calibration,
page 46.
Carry out span calibration,
page 50.
Carry out fresh air calibration,
page 46.
Check DrägerSensor XXS EC4,
page 59.
Repeat function test. Calibrate
or replace DrägerSensor XXS
EC 4, if necessary page 59.
Repeat filter test with X-dock.
Repeat rise time test with X-
dock.
parameters and adjust
Change sensor for applicable
channel with Dräger CC-Vision.
Disconnect and reconnect
power pack or replace the sen-
sor. Sensor must not be loaded
with gas within the first 5 min-
utes.
37