Calling the Info Mode

In measuring mode, press the
If any warning or fault messages exist, the corresponding information or error codes
are displayed (page 30 to page 37).
Press the
key successively for the next display.
OK
The peak values and the exposition values TWA
Warning messages are displayed. Numerical codes of warning messages:
see page 30.
key
OK
Fault messages are displayed. Numerical codes of fault messages: see
page 33.
key
OK
The peak values = the maximum measured values in the case of, e.g., CO,
H
S, ... or the minimum measured values in the case of O
2
interval are displayed
key
OK
The average values of the exposures based on a shift of, e.g., 8 hours (TWA)
of all the active sensors for the exposure evaluation are displayed
key
OK
The short-term values (STEL) = average values of the concentrations over
the average value duration of all the active sensors for the exposure
evaluation are displayed
key
OK
The device is in measuring mode again
If no key is pressed for 10 seconds, the device returns automatically to measuring
mode.

Calling the Info-Off Mode

When the device is in a deactivated state, press the
The name of the gas, measuring unit and measuring range limit value are displayed for
all channels.
Pressing the
key again exits the Info Off mode (or via timeout).
OK
1)
Only when activated in the device configuration. Delivery status: not activated.
key for approx. 3 seconds.
OK
1)
and STEL
1)
are displayed.
within the storage
2
key.