Fault Messages - Drager X-am 5000 Technical Manual

Multi-gas monitor approved as type mqg 0010
Hide thumbs Also See for X-am 5000:
Table of Contents

Advertisement

Special symbol »
and displayed numerical
code:
656
657

Fault messages

Special symbol »
and displayed numerical
code:
102
103
104
105
106
107
108
109
111
112
113
114
115
116
«
Cause
The calibration interval for
DrägerSensor XXS EC 4 has
elapsed
Alarm setpoint A2 setting is
greater than 60 %LEL
«
Cause
The customer's service life
counter has elapsed
The device is defective
Check sum error program
code
The bump test interval has
elapsed
The calibration interval has
elapsed (at least 1 calibration
interval has elapsed)
Bump test error (at least
1 channel has a bump test
error)
The device is defective
The menu function cannot be
carried out because of an
error.
Failed alarm element test:
alarm light.
Failed alarm element test:
alarm horn.
Failed alarm element test:
Vibration motor.
Defective parameter check
Device is disabled by X-dock. Activate device with X-dock.
Failed software update.
Faults, Cause and Remedy
Remedy
Carry out span calibration for
DrägerSensor XXS EC 4,
page 50.
Set alarm setpoint to less than
60 %LEL.
Remedy
Reset the service life counter
using Dräger CC Vision.
The device must be repaired by
DrägerService.
The device must be repaired by
DrägerService.
Carry out bump test, page 43.
Carry out span calibration,
page 47 or page 50.
Carry out bump test, page 43 or
carry out span calibration,
page 47 or page 50.
The device must be repaired by
DrägerService.
Determine the error code via the
info menu and switch it off, if
necessary.
Repeat alarm element test with
Dräger X-dock.
Repeat alarm element test with
X-dock.
Repeat alarm element test with
X-dock.
Correct parameters and repeat
test using X-dock
The device must be repaired by
DrägerService.
33