Configuration
May 2019
14. Click the Service Tools tab to access the device alerts, variables, trends and
74
c) Click the Set Flow Range Limits button and enter a positive value for the
Flow Analysis Lower Velocity Range and the Upper Velocity Range Limits.
When the velocity is outside of the limit parameters, an alert is triggered.
Click the Next button to display the Method Complete page.
maintenance statuses or to edit the configuration parameters.
a) Click the Service Tools → Alerts tab. If an alert condition exists, the alert type
and description displays. Recommended actions are listed to assist you in a
resolution. After you resolve the alert condition, click the Acknowledge
button to clear the alert. Click Apply to write the changes to the meter. If no
alert condition is active, click OK to close the device window.
Figure 4-15: AMS Device Manager - Service Tools Alerts
b) If you change the device configuration, a confirmation dialog displays and
prompts you to write the changes to the meter. Click Yes to write the
changes to the meter or click No to cancel pending changes.
Installation manual
3-9000-771
3818 LNG Liquid Ultrasonic Meter