Electrical Installation
May 2019
3. Check the serial number tag on top of the Transmitter Electronics Enclosure.
4. Pull the wires into the enclosure.
5. Complete the field connection wiring according to the system wiring diagram (see
6. Repeat
7. Apply electrical power to the system to ensure the field connections are working
8. Power down the system and apply the sealing compound to the conduit and allow
9. If required, install the security latches and wire seals on the Transmitter Electronics
10. If required, install the wire seals through the socket head bolts on the Base
11. Connect electrical power to the system.
12. Set or configure the meter using Daniel MeterLink. For additional installation
3.4.2
Startup for systems using flame-proof cable
WARNING
HAZARDOUS VOLTAGE INSIDE
Do not open in flammable gas area. Disconnect all power to the meter.
Failure to comply can result in death or serious injury.
Procedure
1. Check to make certain that all field wiring power is turned OFF.
2. Check the serial number tag on top of the Transmitter Electronics Enclosure.
44
WARNING
HAZARDOUS VOLTAGE INSIDE
Do not open in flammable gas area. Disconnect all power to the meter.
Failure to comply can result in death or serious injury.
Remove the end cap with the serial tag marked with _H1 (nearest its conduit entry)
to gain access to the transmitter electronics.
Appendix
A).
Step 2
through
Step 5
serial tag marked with _H2.
correctly. Replace the end caps and allow the system to run for the time specified
by the customer (usually one week) and an electrician have fully tested the
connections. After the Acceptance Test is witnessed and approved, seal the
conduit.
to set in accordance with manufacturer specifications.
Enclosure end caps (see
Sealing the Transmitter Electronics Enclosure
Figure
3-8).
Enclosure (see
Sealing the Transmitter Electronics
information refer to the system wiring diagram (see
Softeware for Gas and Liquid Ultrasonic Meters Quick Start Manual (P/N 3-9000-763).
Use the Daniel MeterLink Field Setup Wizard to complete the configuration. Also
see,
Field Setup Wizard using Daniel MeterLink
Remove the end cap with the serial tag marked with _H1 (nearest its conduit entry)
to gain access to the transmitter electronics.
for the Transmitter Electronics Enclosure with the
Enclosure).
Appendix
in this manual.
Installation manual
3-9000-771
and see
A), Daniel MeterLink
3818 LNG Liquid Ultrasonic Meter