Risk of minor injuries These installation and operating instructions – NOTICE apply to: Risk of extensive material/property damage V 300 S Order number: 7119-01; 7119-01/002; 7119-02; Other symbols 7119-02/002 These symbols are used in the document and on VS 300 S or in the unit: Order number: 7122-01;...
Important information Safety Order number Dürr Dental has designed and constructed this unit so that when used properly and for the Medical device intended purpose it does not pose any danger to people or property. Health Industry Bar Code (HIBC) Despite this, the following residual risks can remain: Manufacturer...
Important information The unit must not be used as a vacuum Notification requirement of ❯ cleaner. serious incidents Do not use chemicals containing chlorine or ❯ The operator/patient is required to report any foaming chemicals. serious incident that occurs in connection with Operation in operating theatres of explosive ❯...
Important information 2.11 Disposal NOTICE The unit may be contaminated. Instruct Erroneous operation mode due to use the company disposing of the waste to immediately adjacent to other devices take the relevant safety precautions. or with other stacked devices Decontaminate potentially contaminated parts Do not stack the unit together with ❯...
Product description Product description Overview Fig. 1: V 300 S Suction unit Condensation separator Suction hose Exhaust air hose (aluminium) Waste hose LW 20 Set of connection fittings Fig. 2: VS 300 S Combination suction unit Suction hose Exhaust air hose (aluminium)
Product description Technical data V 300 S Electrical data 7119-01 7119-02 Rated voltage 230, 1~ 230, 1~ Mains frequency Nominal current Starting current, approx. 10.4 10.4 Motor protection Motor winding overheat protector 160 °C (±5 °C) Rated power Type of protection...
Product description Ambient conditions during operation Temperature °C +10 to +40 Relative humidity < 70 Classification Medical devices class Electromagnetic compatibility (EMC) Interference emission measurements High-frequency emissions in accordance with CISPR 11 Group 1 Class B Interference voltage at the power supply connection Compliant CISPR 11:2009+A1:2010 Electromagnetic interference radiation...
Page 13
Product description Electromagnetic compatibility (EMC) Interference immunity measurements Immunity to voltage dips, short interruptions and voltage Compliant variations IEC 61000-4-11:2004 Immunity to interference table, near fields of wireless HF communication devices Radio service Frequency band Test level TETRA 400 380 - 390 GMRS 460 430 - 470 FRS 460...
Page 14
Product description Electromagnetic compatibility (EMC) Interference immunity measurements on the supply input Immunity to conducted disturbances, induced by radio- frequency fields – AC mains voltage IEC 61000-4-6:2013 0.15–80 MHz Compliant ISM frequency bands 0.15–80 MHz 80% AM at 1 kHz Immunity to voltage dips, short interruptions and voltage variations Compliant...
Page 15
Product description VS 300 S Electrical data 7122-01 7122-02 7122-03 7122-04 Rated voltage 230, 1~ 230, 1~ 230, 1~ Mains frequency Nominal current Starting current, approx. 10.4 10.4 Motor protection Motor winding overheat protector 160 °C (±5 °C) Rated power Type of protection IP 24 Protection class...
Page 16
Product description Ambient conditions during storage and transport Relative humidity < 95 Ambient conditions during operation Temperature °C +10 to +40 Relative humidity < 70 Classification Medical devices class Electromagnetic compatibility (EMC) Interference emission measurements High-frequency emissions in accordance with CISPR 11 Group 1 Class B Interference voltage at the power supply connection...
Page 17
Product description Electromagnetic compatibility (EMC) Interference immunity measurements Immunity to conducted disturbances, induced by radio- Compliant frequency fields – SIP/SOP ports IEC 61000-4-6:2013 Immunity to power frequency magnetic fields Compliant IEC 61000-4-8:2009 Immunity to voltage dips, short interruptions and voltage Compliant variations IEC 61000-4-11:2004...
Page 18
Product description Electromagnetic compatibility (EMC) Interference immunity measurements on the supply input Immunity to conducted disturbances, induced by radio- frequency fields – AC mains voltage IEC 61000-4-6:2013 0.15–80 MHz Compliant ISM frequency bands 0.15–80 MHz 80% AM at 1 kHz Immunity to voltage dips, short interruptions and voltage variations Compliant...
Page 19
Product description VS 300 S Electrical data 7122-05 Rated voltage 100, 1~ Mains frequency Nominal current Starting current, approx. 20.5 Motor protection Motor winding overheat protector 160 °C (±5 °C) Rated power Type of protection IP 24 Protection class Protective low voltage 24 ~ Output Connections...
Page 20
Product description Ambient conditions during operation Temperature °C +10 to +40 Relative humidity < 70 Classification Medical devices class Electromagnetic compatibility (EMC) Interference emission measurements High-frequency emissions in accordance with CISPR 11 Group 1 Class B Interference voltage at the power supply connection Compliant CISPR 11:2009+A1:2010 Electromagnetic interference radiation...
Page 21
Product description Electromagnetic compatibility (EMC) Interference immunity measurements Immunity to voltage dips, short interruptions and voltage Compliant variations IEC 61000-4-11:2004 Immunity to interference table, near fields of wireless HF communication devices Radio service Frequency band Test level TETRA 400 380 - 390 GMRS 460 430 - 470 FRS 460...
Page 22
Product description Electromagnetic compatibility (EMC) Interference immunity measurements on the supply input Immunity to conducted disturbances, induced by radio- frequency fields – AC mains voltage IEC 61000-4-6:2013 0.15–80 MHz Compliant ISM frequency bands 0.15–80 MHz 80% AM at 1 kHz Immunity to voltage dips, short interruptions and voltage variations Compliant...
Product description Type plate The type plate is is located on the noise reduc- tion hood. Type plate Evaluation of conformity This device has been subjected to conformity acceptance testing in accordance with the cur- rent relevant European Union guidelines. This equipment conforms to all relevant requirements.
Page 24
Product description Operation V 300 S Motor Auxiliary air nozzle Exhaust air muffler Control electronics Suction connection Turbine wheel Exhaust air connection The V-suction unit is suitable for use in dry air suction systems. The advantage of this system is that the suction units can be installed in any suitable room, regardless of the actual connection layout and route- ing of the lines.
Page 25
Product description The exhaust air from the suction unit should be guided out of the building (via the roof where possible). We recommended the installation of a bacteria filter in the exhaust air line. In addition, it is possible to install a noise-reducing muffler in the exhaust air line in order to reduce the amount of noise generated by the unit and by the air flow.
Page 26
Product description VS 300 S Motor Auxiliary air nozzle Exhaust air muffler Control electronics Waste water connection Separation Suction connection Turbine wheel Exhaust air connection The VS suction units are used in wet suction systems. The suction units can be installed on the same floor as the treatment units or on the floor underneath.
Page 27
Product description Inside the separation unit, the aspirated fluids and solid particles pass through a two-stage separation system and are separated from the suction air. This separation system consists of a cyclone separator and a separation turbine. The suction process runs continuously. The aspirated mixture flows into the cyclonic separator, where it is set into a spiral motion.
Assembly Hose materials Assembly For waste connections and suction lines only use the following hose types: – Flexible spiral hoses made of PVC with inte- Requirements grated spiral or equivalent hoses Depending on the suction system, different – Hoses that are resistant to dental disinfectants installation options are available.
Assembly System components Current consumption of Cross-section unit [A] The system components listed below are required or recommended for various procedures > 25 and < 32 or for installation. > 32 and < 40 Rinsing unit > 40 and < 50 It is recommended that the suction system is >...
Assembly > 8% CA 1 Surge tank VS 300 S combination suction unit Bacteria filter For hygienic reasons, we recommend the instal- lation of a bacteria filter in the exhaust air line. If the unit is installed in the surgery and the exhaust air cannot be discharged to the out- doors, it is essential to install a bacteria filter.
❯ Waste water connections must be implemented in accordance with applicable local and national reg- ❯ ulations. V 300 S Hose clamp 25-40 mm Suction hose Æ 30 mm (internal) Waste air pipe (aluminium)Æ 30 mm inside Hose clip Æ 28 mm Hose sleeve Waste water hose Æ...
Page 32
Assembly VS 300 S 10 11 O-ring Æ 30x2 mm Hose connector Æ 30 mm Hose clamp 25-40 mm Waste air pipe (aluminium)Æ 30 mm inside Elbow DN 30 Ring clamp Suction hose Æ 30 mm (internal) Connector Æ 36 mm (external) O-ring Æ...
Assembly 10 Commissioning Electrical connections In many countries technical medical prod- ucts and electrical devices are subject to DANGER regular checks at set intervals. The owner Electric shock due to incorrectly con- must be instructed accordingly. nected device Turn on the unit power switch or the main Never install a mains plug instead of ❯...
Usage 11.3 Once or twice a week before Usage the midday break Under harsher conditions (e.g. hard water 11 Disinfection and cleaning or frequent use of prophy powders) 1x daily before the midday break NOTICE The following are required for cleaning: Device malfunctions or damage due ü...
Usage 12 VS 300 S 12.1 Cleaning the protective strainer WARNING Infection due to contaminated unit Clean and disinfect the suction before ❯ working on the unit. Wear protective equipment when ❯ working (e. g. impermeable gloves, protective goggles and mouth and nose protection).
Page 36
❯ mouth and nose protection). Prior to working on the unit or in case of danger, disconnect it from the mains. 13.1 V 300 S Maintenance interval Maintenance work Every 1-2 years Replace the exhaust air filter (where fitted). * ❯...
Troubleshooting Troubleshooting 14 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. WARNING Infection due to contaminated unit Clean and disinfect the suction before working on the unit. ❯...
Page 38
Troubleshooting Error Possible cause Remedy Suction performance too low Coarse filter blocked Clean the coarse filter at the ❯ intake connection. Leak in the suction line Check and if necessary ❯ establish leak-tightness of suction system and connec- tions. * Mechanical sluggishness of tur- Disassemble the unit and ❯...
Suction connection sealing caps Rinse connection sealing cap Seal all connections with sealing caps. ❯ Water outflow sealing cap Pack the unit securely in preparation for trans- ❯ port. Fig. 3: V 300 S Dummy bushing Securing ring 7119100007L02 1909V001...
Appendix Appendix 16 Handover record This document confirms that a qualified handover of the medical device has taken place and that appropriate instructions have been provided for it. This must be carried out by a qualified adviser for the medical device, who will instruct you in the proper handling and operation of the medical device. Product name Order number (REF) Serial number (SN)