Page 4
Contents 11.3 Once or twice a week before the midday break ....12 Maintenance ..... Troubleshooting 13 Tips for operators and service techni- cians .
These installation and operating instructions apply to: Refer to Operating Instructions. Tyscor V 1 Order number: 7182100200 Tyscor V 2 Wear protective gloves. Order number: 7177-01 Warnings and symbols Disconnect all power from the unit.
Important information Safety This way up / store and transport in an upright position Dürr Dental has designed and constructed this unit so that when used properly and for the intended purpose it does not pose any danger to Keep dry people or property.
Important information The unit must not be used as a vacuum Notification requirement of ❯ cleaner. serious incidents Do not use chemicals containing chlorine or ❯ The operator/patient is required to report any foaming chemicals. serious incident that occurs in connection with Operation in operating theatres of explosive ❯...
Important information Uncontaminated parts (e.g. electronics, plastic ❯ NOTICE and metal parts etc.) should be disposed of in accordance with the local waste disposal regu- Erroneous operation mode due to use lations. immediately adjacent to other devices or with other stacked devices If you have any questions about the correct ❯...
Product description Product description Overview Tyscor V 1 / V 2 suction unit Condensation separator Mains cable with local mains plug Hose LW 50 (0.6 m) Hose LW 50 (1.5 m) Waste water hose LW 20 Set of connection fittings...
Product description Technical data Tyscor V 1 Electrical data * 7182100200 Rated voltage 230, 1~ Mains frequency 50 / 60 Nominal current Rated power Fuses 2x T 4.0 AH / 250 V~ (IEC 60127-2) Type of protection IP 21 Protection class Electrical data are maximum values that can be achieved in the "Boost"...
Product description General data Noise level in accordance with ISO 3746 Network connection LAN technology Ethernet Standard IEEE 802.3u Data rate Mbit/s Connector RJ45 Type of connection Auto MDI-X ³ CAT5 Cable type Ambient conditions during storage and transport Temperature °C -10 to +60 Relative humidity...
Page 13
Product description Electromagnetic compatibility (EMC) Interference immunity measurements Immunity to near fields of wireless HF communication Compliant devices IEC 61000-4-3:2006+A1:2007+A2:2010 Immunity to fast electrical transients/bursts – AC mains Compliant voltage IEC 61000-4-4:2012 Immunity to electrical fast transients/bursts – I/O, Compliant SIP/SOP ports IEC 61000-4-4:2012 Immunity to interference, surges...
Page 14
Product description Immunity to interference levels, near fields of wireless HF communication devices Radio service Frequency band Test level WLAN 802.11 a/n 5100 - 5800 Electromagnetic compatibility (EMC) Interference immunity measurements on the supply input Immunity to fast electrical transients/bursts – AC mains voltage IEC 61000-4-4:2012 Compliant...
Page 15
Product description Electromagnetic compatibility (EMC) Interference immunity measurements SIP/SOP Immunity to conducted disturbances, induced by radio- frequency fields – SIP/SOP ports IEC 61000-4-6:2013 0.15–80 MHz Compliant ISM frequency bands 0.15–80 MHz 80% AM at 1 kHz 7177100004L02 1909V003...
Page 16
Product description Tyscor V 2 Electrical data 7177-01 Rated voltage 230, 1~ Mains frequency 50 / 60 Nominal current 2.7 * Rated power 0.6 * Fuses 2x T 4.0 AH / 250 V~ (IEC 60127-2) Type of protection IP 21 Protection class Maximum values that can be achieved in the "Boost"...
Page 17
Product description Network connection LAN technology Ethernet Standard IEEE 802.3u Data rate Mbit/s Connector RJ45 Type of connection Auto MDI-X Cable type ³ CAT5 Ambient conditions during storage and transport Temperature °C -10 to +60 Relative humidity < 95 Ambient conditions during operation Temperature °C +10 to +40...
Page 18
Product description Electromagnetic compatibility (EMC) Interference immunity measurements Immunity to fast electrical transients/bursts – AC mains Compliant voltage IEC 61000-4-4:2012 Immunity to electrical fast transients/bursts – I/O, Compliant SIP/SOP ports IEC 61000-4-4:2012 Immunity to interference, surges Compliant IEC 61000-4-5:2005 Immunity to conducted disturbances, induced by radio- Compliant frequency fields –...
Page 19
Product description Electromagnetic compatibility (EMC) Interference immunity measurements on the supply input Immunity to fast electrical transients/bursts – AC mains voltage IEC 61000-4-4:2012 Compliant ± 2 kV 100 kHz repetition rate Immunity to surges, line-to-line IEC 61000-4-5:2005 Compliant ± 0.5 kV, ± 1 kV Immunity to surges, line-earth IEC 61000-4-5:2005 Compliant...
Product description Type plate The type plate can be found on the housing upper part. Type plate Evaluation of conformity This device has been subjected to conformity acceptance testing in accordance with the cur- rent relevant European Union guidelines. This equipment conforms to all relevant requirements.
Page 21
Product description Operation Suction connection CAN bus Network connection Mains connection with mains fuses Control connection Condensation drain Exhaust air connection Yellow LED – Ethernet Red LED - radial blower fault 10 Green LED – ready for operation 11 Blue LED – start signal 12 Manual start button 13 SD card slot (Micro SD) The V suction unit is used in "dry"...
Page 22
Operating modes Tyscor V 1 With Tyscor V/VS 1 there is one operating mode available. The Tyscor V/VS 1 can be used to supply one treatment chair. With the aid of a special upgrade, the performance can be raised to the level of a Tyscor V/VS 2.
Assembly Hose materials Assembly For waste connections and suction lines only use the following hose types: – Flexible spiral hoses made of PVC with inte- Requirements grated spiral or equivalent hoses The unit can be installed on the same level as the –...
Assembly System components Installation The system components listed below are The actual connection can vary depend- required or recommended for various procedures ing on the chosen installation option. The or for installation. connection shown is only an example. Exhaust air filter / bacteria fil- Mounting the condensate separator For reasons of hygiene, we recommend the...
Assembly Installation and routeing of hoses and pipes Establish connections between the pipe system and the unit using the flexible hoses supplied. This ❯ will prevent vibrations from being transmitted to the pipe system. The connection between the pipe line and unit suction connection should be kept as short as possi- ❯...
Page 26
Assembly Network connection Purpose of the network connection The network connection is used to exchange information or control signals between the unit and a software installed on a computer, in order to, e. g.: – Display parameters – Select operating modes –...
Page 27
Assembly PCB (main board) electrical connections Ethernet/CAN X11 X9 Control voltage output, 24 V DC, 400 mA, control signal input Network connection and CAN bus Supply voltage 230 V Motor control fan connection 2 Service interface SD card holder (for Micro SD) Separation motor RPM monitor (VS only) Separation motor supply voltage (VS only) Motor control fan connection 1...
Assembly Commissioning Upgrade/update In many countries technical medical prod- With the aid of a Micro SD card it is possible to ucts and electrical devices are subject to perform an upgrade/update on the unit. regular checks at set intervals. The owner Do not disconnect the unit from the mains must be instructed accordingly.
Page 29
Assembly When setting up the PC system in the vicinity ❯ of the patients: Only connect components (e.g. computer, monitor, printer) that comply with the standard IEC 60601‑1 (EN 60601‑1). When setting up the PC system outside of the ❯ vicinity of the patients: Connect components (e.g.
Page 30
Usage 10.5 Fault Usage RED LED illuminates A fault is present in the radial blower or in 10 LEDs the separation system (VS only). Yellow LED – network connected Red LED – fault in the radial blower / sepa- ration system (VS only) Green LED –...
Usage 11 Disinfection and cleaning 11.3 Once or twice a week before the midday break NOTICE Under harsher conditions (e.g. hard water Device malfunctions or damage due or frequent use of prophy powders) 1x to use of incorrect media daily before the midday break Guarantee claims may become invalid The following are required for cleaning: as a result.
Usage 12 Maintenance All maintenance work must be performed by a qualified expert or by one of our Service Techni- cians. WARNING Infection due to contaminated unit Clean and disinfect the suction before working on the unit. ❯ Wear protective equipment when working (e. g. impermeable gloves, protective goggles and ❯...
Troubleshooting Troubleshooting 13 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. WARNING Infection due to contaminated unit Clean and disinfect the suction before working on the unit. ❯...
Troubleshooting Error Possible cause Remedy Suction performance too low Leak in the suction pipe Check and if necessary ❯ establish leak-tightness of suction pipe and connections. Poor pipe routeing Use higher operating mode ❯ level. Membrane valve defective Check the membrane valve at ❯...
Troubleshooting Error Possible cause Remedy Firmware mismatch Different firmware versions on Perform/repeat the firmware ❯ the two processors after a update. * firmware update Speed Feedback Failure Motor speed detection defective Replace the suction stage. * ❯ MC Lib failure Control error Replace the electronics.
Troubleshooting 14 Transporting the unit WARNING Infection due to contaminated unit Disinfect the unit before transport. ❯ Close all media connections. ❯ Wear protective equipment to avoid any risk of infection (e.g. liquid-tight protective gloves, protective goggles, face mask). Before disassembly, clean and disinfect the ❯...
Appendix Appendix 15 Handover record This document confirms that a qualified handover of the medical device has taken place and that appropriate instructions have been provided for it. This must be carried out by a qualified adviser for the medical device, who will instruct you in the proper handling and operation of the medical device. Product name Order number (REF) Serial number (SN)
Page 40
Hersteller/Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...