These operating instructions are valid for Observe the operating instructions. Power Tower View, order number: A949444000, A949443000, A949454000, A949453000, CE labelling with the number of the noti- A949242000, A949252000 and A949243000.
Important information Safety Disconnect all power from the unit. Dürr Dental has designed and constructed this unit so that when used properly and for the inten- ded purpose it does not pose any danger to Monitor ambient conditions people or property. Despite this, the following residual risks can remain: Copyright information...
Important information Improper use The following groups are not permitted to operate or use a commercially operated unit: Unit – People without the necessary experience and Any other usage or usage beyond this scope is knowledge deemed to be improper. –...
Important information Keep the packing materials out of the reach of ❯ children. 2.10 Disposal The unit must be disposed of properly. Within the European Union, the unit must be disposed of in accordance with EU Directive 2012/19/EU (WEEE). If you have any questions about the correct ❯...
– OroCup Care system delivery (possible variations due to country-spe- Power Tower View ....A949252000 cific requirements and/or import regulations): Power Tower View ....A949444000 –...
Page 10
Product description To configure the required filters or filter sets, you can also use our filter configura- tor at: www.duerrdental.com/filterkonfigurator Any repairs exceeding routine mainte- nance may only be carried out by quali- fied personnel or our service. Information about replacement parts is available from the portal for authorised specialist dealers at: www.duerrdental.net...
Product description Technical data Electrical data Frequency Protection class Type of protection IP 20 Operating mode General technical data Dimensions (W x H x D) 640 x 2080 x 610 Pressure tank volume Pressure dew point at 7 bar (0.7 MPa) * °C <5 Safety valve settings, maximum permitted...
Product description Ambient conditions during storage and transport Relative humidity Max. 95 Ambient conditions during operation Temperature °C +10 to +40 Relative humidity Max. 70 Height above sea level < 2000 400 V 3~ variant Delivery at 5 bar (0.5 MPa)* l/min Noise level ** dB(A)
Page 13
Product description Max. permitted mains impedance in 0.24 Ω accordance with IEC 61000-3-11 1 compressor unit without drying unit At an ambient temperature of +40 °C, noise level in accordance with ISO 3744 At an ambient temperature of +35 °C, noise level in accordance with ISO 3744 PTV A949242000 with second Duo unit General technical data Weight...
Product description Type plate The type plate is located at the bottom right of the housing. Declaration of conformity This device has been subjected to conformity acceptance testing in accordance with the cur- rent relevant European Union guidelines. This equipment conforms to all relevant requirements. A949100052L02 2206V006...
Product description Operation Dry suction system Wet suction system Main power switch Drying unit Suction unit Main power switch Condensation separator Drying unit Compressor unit Suction unit Pressure tank Pressure equalisation tank Touch display Compressor unit Pressure tank Touch display A949100052L02 2206V006...
Product description Functional description The Power Tower View combines compressed air generation, suction and amalgam separation in one housing. The housing is insulated, ensuring an appropriate level of soundproofing. The unit is operated via a touch display and indicates the various parameters of the individual components.
Assembly Pipe materials Assembly Only use vacuum-sealed HT-waste pipes manufactured from the following materials: – Polypropylene (PP), Only qualified specialists or employees – Chlorinated polyvinyl chloride (PVC-C), trained by Dürr Dental are permitted to – Plasticizer-free polyvinyl chloride (PVC-U), install, connect and start using the unit. –...
Assembly The connection to the mains supply must be a ❯ Installation type Line layout (minimum fixed connection that cannot be disconnected requirements) without the use of tools. Plug-in connections Flexible – PVC data cable with (power outlet/plug) are not permissible. shielded cable sheath- Observe the current consumption of the devi- ❯...
Assembly Installation Install the compressed air module Combining devices safely Mount the compressed air module on the suc- ❯ Take care when connecting units together or to tion module using four M5x20 thread-cutting parts of other systems as there is always an ele- screws.
Assembly Push on and secure the noise insulation on the Position the unit over the foam in such a way ❯ ❯ suction manifold of the compressor unit. that the connections are easily accessible. Positioning the unit Place the foam over the connections in the ❯...
Assembly Align the unit in its final position. Installing a second compres- ❯ sor unit (optional) CAUTION Functional disorder in case of incor- rect installation Unit does not start up. When retrofitting a compressor unit in ❯ the case of 400 V variants, the com- pressor unit must have a star connec- tion.
Assembly Install the drying unit Connect the drying unit to the compressed air ❯ hose coming out of the cooler. Install the drying unit. ❯ Insert nozzle, see installation instructions ❯ 9000-610-49/01. Insert the filter and close the drying unit with the cover. A949100052L02 2206V006...
Assembly Connect the drying unit to the pressure tank. Additional work for 230 V var- ❯ iant Operation with two compressor units Remove the earth strap. ❯ Remove the clip for L bracket. ❯ Mount the valve block with earth strap. ❯...
Page 24
Assembly Establish the compressed air connection Lay the valve cable from the valve block to the ❯ ❯ between the compressor unit and valve block. PCB. Mount the second compressor unit and estab- ❯ lish the compressed air connection. Affix a spacer between the compressed air ❯...
Assembly Operation with one compressor unit Install the amalgam separator Mount the L bracket with the compressed air ❯ (optional) hose to the cooler. Position the amalgam separator in the lower ❯ part of the cabinet. Remove the electronic hood. ❯...
Velcro and connect to the control board. Position the amalgam separator correctly in the ❯ Power Tower View. Plug the network cable into the network socket ❯ on the BUS adapter and connect to the RJ45 at X31 on the PCB of the device.
Assembly Commissioning MAIN Perform an electrical safety check prior to commissioning. Switch on the main power switch. ❯ The compressor will start and switches off at approx. 0.75 MPa. In the case of a 230 V variant with a compres- ❯...
Assembly Checking the switch-on/cut- Checking the safety valve off pressure Correct operation of the safety valve must be checked when the unit is started up for the first The switch-on/cut-off pressure is preset at the time and again subsequently at regular intervals. factory.
Assembly An adequate ventilation by fans in the unit Port Purpose Service is only possible when all covers are 502 TCP Device data attached. Do not cover the air inlets on Event log data Syslog the housing. 22 TCP, 23 Diagnosis Telnet, Place all covers on the unit.
Usage to switch to the detail view for the ❯ Usage alarm texts. Calling up messages on the Operating the touch touch screen screen Messages are divided into the following catego- ries: NOTICE Fault Unit will no longer function. When the error has been Damage to the touch screen due to remedied, it may be neces- incorrect handling...
Usage 10 Operation 10.1 Switch on the unit. Switch on the unit at the main power switch. ❯ 10.2 Device information Use the menus Unit information > Unit data and Unit information > Unit usage data to call up information including: –...
Usage 11 Maintenance Prior to working on the unit or in case of danger, disconnect it from the mains. Comply with the specifications of the Installation and Operating Instructions for the devices installed. ❯ 11.1 Compressor unit and drying unit WARNING Risk of infection due to burst filters Particles enter the compressed air network and can therefore enter the mouth of the patient.
Page 33
Usage Replacing the air intake filter Switch off the device. ❯ Pull off the noise reducer from the air intake fil- ❯ ter. Remove the air intake filter. ❯ Insert a new air intake filter. ❯ Push on the noise reducer onto the air intake ❯...
Troubleshooting Troubleshooting 12 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. Prior to working on the unit or in case of danger, disconnect it from the mains. Error Possible cause Remedy...
Appendix 14 Electrical connection plan 14.1 Version 400 V, 3N~ MAIN MAC- Adresse L1 L2 L3 N PE Alarm max. 60 V, 400 mA PE U V W PE U V W PE U V W L N PE A949100052L02 2206V006...
Appendix 14.2 Version 230 V, 1N~ MAIN Barcode Valve2 Valve1 MAC- Adresse L N PE Alarm max. 60 V, 400 mA PE U V W L N PE L N PE L N PE A949100052L02 2206V006...
Appendix 15 Menu structure 15.1 Settings Level 1 Level 2 Level 3 Level 4 Operator Access level Administrator Service Technician (PIN) Device data Device Information Device usage data German (DE) System settings Language English (EN) ... Time zone Date / time Date Time DHCP...
Page 39
Appendix Visible from access level Operator or higher Visible from access level Administrator or higher Visible from access level Service Technician or higher The access level "Service Technician" offers additional functions that must only be adjusted by Dürr Dental or by individuals/companies authorised to do so by Dürr Dental. Corresponding documents for this can be downloaded from www.duerrdental.net.
Appendix 16 Handover record This document confirms that a qualified handover of the medical device has taken place and that appropriate instructions have been provided for it. This must be carried out by a qualified adviser for the medical device, who will instruct you in the proper handling and operation of the medical device. Product name Order number (REF) Serial number (SN)
Page 42
Hersteller / Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...
Need help?
Do you have a question about the Power Tower View and is the answer not in the manual?
Questions and answers