Description:
Action:
50
LATITUDE
The Communicator was unable to
start an interrogation of your implanted
device, or your implanted device
was out of range at the time of the
attempted interrogation.
• Ensure the Communicator is
optimally placed as described
in "Where to Place Your
Communicator" on page 14.
• Face the Communicator. Sit directly
in front of the Communicator. Make
sure you are within 1 meter (3 feet)
of the Communicator.
• Move any wireless electronic
products (such as cordless or
cellular phones or baby monitors)
that are within 1 meter (3 feet) of the
Communicator.
To verify that troubleshooting was a
success:
• Press the Heart button to start
another interrogation. If the
interrogation was a success, all
three Collecting Waves will light
green for 2 minutes.
Communicator Patient Manual